SUCTION 3724222 HOUSE IRRIGATOR 5X7FR

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2013-04-22 for SUCTION 3724222 HOUSE IRRIGATOR 5X7FR manufactured by Medtronic Xomed, Inc..

Event Text Entries

[3431423] It was reported there was an issue with four instruments at a cadaver lab in the same day. The head of the irrigation port is broke off of two of the instruments. The other two are bent. This happened during use of the instruments. There were no patients involved in the case, used in a cadaver lab.
Patient Sequence No: 1, Text Type: D, B5


[10779750] Blank fields on this report are the result of information not being provided by initial reporter. This device is used for therapeutic purposes. Part # 3724225, suction house irrigator 8x12fr, qty of 2, lot #, serial #, and manufacture date unknown. (b)(4). The quality engineer reviewed the returned instruments and concluded that the damage is most likely due to excessive force from the end user. With the exception of the bends or breaks defined in the complaint, the remainder of the instruments is intact and no visible damage is seen. The remainders of the instruments meet specifications. The irrigation ports are attached to stainless steel tubing, which in itself provides good support and does not bend easily. One irrigation port had the weld break and it came off the stainless steel tubing. The second unit had both the weld and the stainless steel tubing break off, which support the conclusion of excessive force being used. Results - stress problem (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1045254-2013-00343
MDR Report Key3071103
Report Source06
Date Received2013-04-22
Date of Report2013-03-29
Date Mfgr Received2013-03-29
Date Added to Maude2013-06-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactCHRISTY CAIN
Manufacturer Street6743 SOUTHPOINT DRIVE NORTH
Manufacturer CityJACKSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal32216
Manufacturer Phone9043328353
Manufacturer G1MEDTRONIC XOMED, INC.
Manufacturer Street6743 SOUTHPOINT DRIVE NORTH
Manufacturer CityJACKSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal Code32216
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSUCTION 3724222 HOUSE IRRIGATOR 5X7FR
Generic NameTUBE, EAR SUCTION
Product CodeJZF
Date Received2013-04-22
Returned To Mfg2013-04-09
Model Number3724222
Catalog Number3724222
Lot NumberNOT PROVIDED
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC XOMED, INC.
Manufacturer Address6743 SOUTHPOINT DRIVE NORTH JACKSONVILLE FL 32216 US 32216


Patients

Patient NumberTreatmentOutcomeDate
10 2013-04-22