NEO MAPCATH SENSOR STYLET 0.8FR 60-2008

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2013-04-15 for NEO MAPCATH SENSOR STYLET 0.8FR 60-2008 manufactured by Corpak Medsystems.

Event Text Entries

[3349801] Event description: the stylet used in advancing the picc line broke while inside the pt and became an intravascular foreign body. The piece required surgical intervention (cardiac catheterization) for removal. The "piece" was removed without incident. According to the customer, the stylet was inserted into the picc prior to insertion. There was no reported resistance upon insertion or removal. The picc line was shortened prior to placing stylet. According to the customer, the picc was not flushed prior to stylet removal.
Patient Sequence No: 1, Text Type: D, B5


[10779803] There was no report of pt injury. The actual sample has not been returned. Samples from the same lot were tested at greater than 17n which is greater than any force expected in the clinical setting. No mfg or design defect was identified in the investigation. The clinician did not report any difficulties with stylet insertion or picc placement.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3009124963-2013-00011
MDR Report Key3075015
Report Source05,06
Date Received2013-04-15
Date of Report2013-04-09
Date of Event2013-02-05
Date Mfgr Received2013-03-22
Device Manufacturer Date2011-05-01
Date Added to Maude2013-04-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer Street1001 ASBURY DR
Manufacturer CityBUFFALO GROVE IL 60089
Manufacturer CountryUS
Manufacturer Postal60089
Manufacturer Phone8474033400
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameNEO MAPCATH SENSOR STYLET
Generic NameNAVIGATOR WIRE
Product CodeOMF
Date Received2013-04-15
Model Number0.8FR
Catalog Number60-2008
Lot Number4427
Device Expiration Date2016-03-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerCORPAK MEDSYSTEMS
Manufacturer Address1001 ASBURY DR BUFFALO GROVE IL 60089 US 60089


Patients

Patient NumberTreatmentOutcomeDate
101. Other; 2. Required No Informationntervention 2013-04-15