DIMENSION EXL

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2013-05-08 for DIMENSION EXL manufactured by Siemens Healthcare Diagnostics.

Event Text Entries

[19196523] Discordant sodium (na) and potassium (k) results were obtained on a dimension exl instrument for three patients. The initial results were not reported to the physicians. The samples were retested and the repeated results were reported to the physicians. There are no known reports of patient intervention or adverse health consequences due to the discordant sodium and potassium results.
Patient Sequence No: 1, Text Type: D, B5


[19471444] A siemens field service engineer (fse) was dispatched to the customer site and analyzed the instrument and instrument data. The fse determined that the cause of the discordant sodium and potassium results were due to user error; improper sample tube spinning. The tube vendor recommends pulling 100 - 1300 g force for 10 minutes in order to get a clean sample. Siemens instructed the customer to follow the tube manufacturer's recommendation for proper centrifugation times and speeds. The instrument is performing within specifications and further evaluation of the device is not required.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1226181-2013-00210
MDR Report Key3100871
Report Source05,06
Date Received2013-05-08
Date of Report2013-04-09
Date of Event2013-04-05
Date Mfgr Received2013-04-09
Device Manufacturer Date2008-05-21
Date Added to Maude2013-07-12
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactJOHN NELSON
Manufacturer Street511 BENEDICT AVENUE
Manufacturer CityTARRYTOWN NY 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145242530
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS INC- BROOKFIELD
Manufacturer Street101 SILVERMINE ROAD
Manufacturer CityBROOKFIELD CT 06804
Manufacturer CountryUS
Manufacturer Postal Code06804
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDIMENSION EXL
Generic NameCLINICAL CHEMISTRY SYSTEM
Product CodeNGS
Date Received2013-05-08
Model NumberDIMENSION EXL
OperatorMEDICAL TECHNOLOGIST
Device AvailabilityN
Device Age5 YR
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS
Manufacturer Address101 SILVERMINE ROAD BROOKFIELD CT 06804 US 06804


Patients

Patient NumberTreatmentOutcomeDate
10 2013-05-08

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.