MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2013-05-20 for WAVESCAN WAVEFRONT SYSTEM WS1 0070-0150 manufactured by Abbott Medical Optics.
[3500008]
The clinic reported that the system gave off a burning smell and smoke was observed coming from the computer.
Patient Sequence No: 1, Text Type: D, B5
[10843901]
The field service engineer evaluated the system at the customer location and found that the power supply had a burnt odor. The power supply was replaced and the system was tested and placed back into service. All pertinent information available to amo has been submitted.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 3006695864-2013-00168 |
MDR Report Key | 3120850 |
Report Source | 05,06 |
Date Received | 2013-05-20 |
Date of Report | 2013-04-25 |
Date of Event | 2013-04-25 |
Date Mfgr Received | 2013-04-25 |
Device Manufacturer Date | 2003-02-25 |
Date Added to Maude | 2013-05-20 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 0 |
Event Location | 0 |
Manufacturer Contact | CAROL KAIL |
Manufacturer Street | 1700 EAST ST. ANDREW PLACE |
Manufacturer City | SANTA ANA CA 92705 |
Manufacturer Country | US |
Manufacturer Postal | 92705 |
Manufacturer Phone | 7142478598 |
Manufacturer G1 | ABBOTT MEDICAL OPTICS INC. |
Manufacturer Street | AMO MILPITAS 510 COTTONWOOD DRIVE |
Manufacturer City | MILPITAS CA 95035 |
Manufacturer Country | US |
Manufacturer Postal Code | 95035 |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | WAVESCAN WAVEFRONT SYSTEM WS1 |
Generic Name | REFRACTIVE MEASUREMENT |
Product Code | HKO |
Date Received | 2013-05-20 |
Model Number | WS1 |
Catalog Number | 0070-0150 |
Operator | OTHER HEALTH CARE PROFESSIONAL |
Device Availability | Y |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | ABBOTT MEDICAL OPTICS |
Manufacturer Address | SANTA ANA CA US |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2013-05-20 |