WAVESCAN WAVEFRONT SYSTEM WS1 0070-0150

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2013-05-20 for WAVESCAN WAVEFRONT SYSTEM WS1 0070-0150 manufactured by Abbott Medical Optics.

Event Text Entries

[3500008] The clinic reported that the system gave off a burning smell and smoke was observed coming from the computer.
Patient Sequence No: 1, Text Type: D, B5


[10843901] The field service engineer evaluated the system at the customer location and found that the power supply had a burnt odor. The power supply was replaced and the system was tested and placed back into service. All pertinent information available to amo has been submitted.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3006695864-2013-00168
MDR Report Key3120850
Report Source05,06
Date Received2013-05-20
Date of Report2013-04-25
Date of Event2013-04-25
Date Mfgr Received2013-04-25
Device Manufacturer Date2003-02-25
Date Added to Maude2013-05-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactCAROL KAIL
Manufacturer Street1700 EAST ST. ANDREW PLACE
Manufacturer CitySANTA ANA CA 92705
Manufacturer CountryUS
Manufacturer Postal92705
Manufacturer Phone7142478598
Manufacturer G1ABBOTT MEDICAL OPTICS INC.
Manufacturer StreetAMO MILPITAS 510 COTTONWOOD DRIVE
Manufacturer CityMILPITAS CA 95035
Manufacturer CountryUS
Manufacturer Postal Code95035
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameWAVESCAN WAVEFRONT SYSTEM WS1
Generic NameREFRACTIVE MEASUREMENT
Product CodeHKO
Date Received2013-05-20
Model NumberWS1
Catalog Number0070-0150
OperatorOTHER HEALTH CARE PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerABBOTT MEDICAL OPTICS
Manufacturer AddressSANTA ANA CA US


Patients

Patient NumberTreatmentOutcomeDate
10 2013-05-20

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.