SUPPORT ARM 177

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,06,07 report with the FDA on 2013-05-16 for SUPPORT ARM 177 manufactured by Maquet Critical Care Ab.

Event Text Entries

[18249477] (b)(4).
Patient Sequence No: 1, Text Type: D, B5


[18324029] Only part of the support arm was returned. The part that was returned was the lowest pivot arm and it was broken in the articulating join. Visual inspection showed that the broken section was porous. The returned part did not include date stamp but according to the customer it was purchased in 2005. The customer stated that support arms are not overloaded or used to pull ventilators during transport but stated that sometimes the support arms "touch" the large stands. This can imply that impacts occur. The support arm is a casting and it most probably developed a crack at an earlier occasion either by overloading or impact and this led to the reported breaking. Previous investigations led to a redesign and a change of mfg process in order to obtain a higher mechanical strength of the support arms. This change was implemented in production during (b)(6) 2009. The support arm in this report was manufactured before implementation of this change. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number8010042-2013-00076
MDR Report Key3126589
Report Source01,05,06,07
Date Received2013-05-16
Date Mfgr Received2013-04-18
Device Manufacturer Date2005-01-01
Date Added to Maude2013-05-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional0
Initial Report to FDA0
Report to FDA0
Event Location0
Manufacturer ContactJANICE PEVIDE
Manufacturer Street45 BARBOUR POND DR
Manufacturer CityWAYNE NJ 07470
Manufacturer CountryUS
Manufacturer Postal07470
Manufacturer Phone9737097753
Manufacturer G1MAQUET CRITICAL CARE AB
Manufacturer CitySOLNA S-17154
Manufacturer CountrySW
Manufacturer Postal CodeS-17154
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSUPPORT ARM 177
Product CodeIOY
Date Received2013-05-16
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMAQUET CRITICAL CARE AB
Manufacturer Address2 ROENTGENVAGEN SOLNA SW


Patients

Patient NumberTreatmentOutcomeDate
10 2013-05-16

© 2026 FDA.report