RUSCH GREENSPEC FO HANDLE, MEDIUM 004411100

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2013-05-20 for RUSCH GREENSPEC FO HANDLE, MEDIUM 004411100 manufactured by Teleflex.

Event Text Entries

[3564258] The event is reported as: the biomed technician states that the handle would heat up if the laryngoscope blade was attached to the handle and left on. Biomed noticed this during maintenance that the insulator ring in the head of the handle was either damaged or missing. No pt injury/involvement reported.
Patient Sequence No: 1, Text Type: D, B5


[10744847] The device sample was not returned for eval at the time of this report.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1044475-2013-00074
MDR Report Key3127285
Report Source06
Date Received2013-05-20
Date of Report2013-04-30
Date of Event2013-04-24
Date Mfgr Received2013-04-30
Date Added to Maude2013-07-26
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMARGIE BURTON, RN
Manufacturer StreetPO BOX 12600
Manufacturer CityRESEARCH TRIANGLE PARK NC 27709
Manufacturer CountryUS
Manufacturer Postal27709
Manufacturer Phone9194334965
Manufacturer G1TELEFLEX
Manufacturer Street2917 WECK DR.
Manufacturer CityRESEARCH TRIANGLE PARK NC 27709
Manufacturer CountryUS
Manufacturer Postal Code27709
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRUSCH GREENSPEC FO HANDLE, MEDIUM
Generic NameFIBEROPTIC LARYNGOSCOPE HANDLE
Product CodeEQN
Date Received2013-05-20
Catalog Number004411100
Device AvailabilityY
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerTELEFLEX
Manufacturer AddressRESEARCH TRIANGLE PARK NC 27709 US 27709


Patients

Patient NumberTreatmentOutcomeDate
10 2013-05-20

© 2026 FDA.report