OLYMPIC COOL-CAP SYSTEM 60010

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2013-05-21 for OLYMPIC COOL-CAP SYSTEM 60010 manufactured by Olympic Medical Incorporated.

Event Text Entries

[18055689] An infant was being treated for hypoxic ischemic encephalopathy with a cool-cap system when the screen froze twice. It was reported that the device gave no alarms or messages either time the screen froze. The system was rebooted both times and treatment resumed. There was no reported pt harm. Device will be returned for eval. (b)(4).
Patient Sequence No: 1, Text Type: D, B5


[18324417] To be returned. The malfunction described in this event is the cause of an ongoing field correction recall (fda recall #z-0447-2013, natus capa (b)(4)). Natus is submitting a pma supplement for approval of a software upgrade that addresses this issue. There will be no follow-up to this report unless natus' review of this device shows a different failure/mode mechanism than described in the complaint.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3018859-2013-00003
MDR Report Key3127438
Report Source05
Date Received2013-05-21
Date of Report2013-05-17
Date of Event2013-04-26
Device Manufacturer Date2009-02-26
Date Added to Maude2013-07-30
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactSTEVE HESLER
Manufacturer Street5900 FIRST AVE., SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal98108
Manufacturer Phone2062685156
Single Use3
Remedial ActionRC
Previous Use Code3
Removal Correction NumberZ-0447-2013
Event Type3
Type of Report3

Device Details

Brand NameOLYMPIC COOL-CAP SYSTEM
Generic NameCOOLING CAP (INFANTS)
Product CodeMXM
Date Received2013-05-21
Model Number60010
Catalog Number60010
Lot NumberNA
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerOLYMPIC MEDICAL INCORPORATED
Manufacturer AddressSEATTLE WA US


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2013-05-21

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.