BELIMED

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,07 report with the FDA on 2008-07-29 for BELIMED manufactured by Belimed Ag.

Event Text Entries

[19726192] (b)(4).
Patient Sequence No: 1, Text Type: D, B5


[19833028] Belimed technician evaluated the unit and found water leaking from the wash pump, either the shaft seals or casings, and from various hoses under the machine. Also noted were signs of corrosion and mold under the machine. The washer/disinfector is not maintained by belimed personnel.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number8043641-2008-00001
MDR Report Key3131174
Report Source05,07
Date Received2008-07-29
Date Mfgr Received2008-07-17
Device Manufacturer Date2000-06-01
Date Added to Maude2013-05-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional0
Initial Report to FDA0
Report to FDA0
Event Location0
Manufacturer Street2284 CLEMENTS FERRY ROAD
Manufacturer CityCHARLESTON SC 28482
Manufacturer CountryUS
Manufacturer Postal28482
Manufacturer Phone8432167424
Manufacturer G1BELIMED AG
Manufacturer StreetDORFSTRASSE 4
Manufacturer CityBALLWIL 6275
Manufacturer CountrySZ
Manufacturer Postal Code6275
Single Use3
Remedial ActionRP
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameBELIMED
Generic NameWASHER/DISINFECTOR
Product CodeFLH
Date Received2008-07-29
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBELIMED AG
Manufacturer AddressBALLWIL SZ


Patients

Patient NumberTreatmentOutcomeDate
10 2008-07-29

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.