DC-II COLLINS FILTER K022454

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 00,05,06 report with the FDA on 2006-04-27 for DC-II COLLINS FILTER K022454 manufactured by Ferraris Respiratory Inc.

Event Text Entries

[3454200] Event description as reported by user through medwatch program: during diffusion test, pt inhaled fast and deep and started to cough uncontrollably. Pt coughed up a small (approximately 3/4" x 1/2 cm) piece of plastic which came from the inside of the pt filter on the mouthpiece. Pulmonary function test continued with different mouthpiece and filter. Device usage problem: device failed (e. G. Broke, couldn't get it to work or stopped working).
Patient Sequence No: 1, Text Type: D, B5


[10739891] Suspect that plastic described by user is from a damaged filter support fin the broke from one of the plastic halves. First report of such issue. Have reviewed filter assembly work instructions that note that inspection of components is to take place prior to the sonic welding step. Will edit the instructions with pictures to identify specific issues/areas to be inspected and unacceptable conditions to be looking for.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1722076-2006-00001
MDR Report Key3131333
Report Source00,05,06
Date Received2006-04-27
Date of Report2006-03-28
Date of Event2006-02-10
Date Mfgr Received2006-03-28
Device Manufacturer Date2006-05-01
Date Added to Maude2013-05-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactDON HENTON
Manufacturer Street901 FRONT ST
Manufacturer CityLOUISVILLE CO 80027
Manufacturer CountryUS
Manufacturer Postal80027
Manufacturer G1FERRARIS RESPIRATORY INC
Manufacturer Street1301 COURTESY RD
Manufacturer CityLOUISVILLE CO 80027
Manufacturer CountryUS
Manufacturer Postal Code80027
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDC-II COLLINS FILTER
Generic NameMOUTHPIECE FILTER FOR PULMONARY FUNCTION TEST
Product CodeBZC
Date Received2006-04-27
Model NumberK022454
Catalog NumberK022454
Lot Number051705
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerFERRARIS RESPIRATORY INC
Manufacturer Address901 FRONT ST LOUISVILLE CO 80027 US 80027


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2006-04-27