TONOMETER

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 08 report with the FDA on 2013-05-09 for TONOMETER manufactured by .

Event Text Entries

[21947775] No complaints of this nature have been received from any other customers. It is clear from the information given by the users that the cleaning and disinfection process being used differs from the keeler instructions in a number of areas: only the tip is immersed, prism face down in the solution. The prisms are not adequately rinsed prior to drying; the prisms were not inspected for cracks or chips prior to use on the patient(s). From the testing conducted at keeler during this investigation we have concluded that processing the keeler applanation tonometer prisms using 3% hydrogen peroxide solution, and following the user instructions will have no detrimental effect on the prism, and is therefore deemed safe and appropriate for use. The damage to the prism surfaces and subsequent reversible injures sustain by the patients have been caused by a combination of improper cleaning and disinfection techniques and a failure on the part of the user to ensure the condition of the prism was appropriate for continued use as specified in the user instructions.
Patient Sequence No: 1, Text Type: N, H10


[22042565] Leaking or cracked tonometer prism tip takes up small amounts of disinfectant solution while cleaning, which is then released onto the patients eye when next used. This solution irritates the patients eye or causes reversible corneal scarring. Reporter's own words; "i recently got a tonometry tip and i have noticed that it leaks after i clean it. I have already left scars on two patients already and i am afraid of burning more patients. I need to be able to use a tonometry tip everyday so i was wondering if i could get a new one if i gave mine back. " "hi, i am writing in regards to my tonometer tip. I came in and spoke about it a few weeks ago now. Sorry for the delayed email but my tip has cracks in it and this is after proper use of the tip. It was correctly cleaned, i only soaked it in peroxide for the required 10 minutes and it was properly inserted, i never forced it onto the slit lamp. I would like a replacement because this is imperative for my clinical experience. " "unfortunately, i am having some issues with my tonometry tip. Over the past few weeks i've begun to notice progressive cracks on the applanation end of my tonometer. I understand that these cracks can lead to "tatoo-ing" of the patient's cornea as well as make tonometry readings difficult. The cleaning technique that i have been using is an 10 minute soak in 3% peroxide solution. Please let me know what my options for replacement are. "
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1000391004-2013-00001
MDR Report Key3131482
Report Source08
Date Received2013-05-09
Date of Report2013-04-19
Date Mfgr Received2013-04-18
Device Manufacturer Date2012-09-01
Date Added to Maude2013-07-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer StreetCLEWER HILL RD.
Manufacturer CityWINDSOR, BERKSHIRE SL4 4AA
Manufacturer CountryUK
Manufacturer PostalSL4 4AA
Manufacturer Phone1753827153
Single Use3
Remedial ActionRL
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTONOMETER
Generic NameTONOMETER
Product CodeHKY
Date Received2013-05-09
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0


Patients

Patient NumberTreatmentOutcomeDate
10 2013-05-09

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.