LINEAR SERIES COLLIMATORS LINEAR MC150, LINEAR II, LINEAR III, LINEAR IV 70-63000, 70-43000, 70-53000,

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 08 report with the FDA on 2001-01-29 for LINEAR SERIES COLLIMATORS LINEAR MC150, LINEAR II, LINEAR III, LINEAR IV 70-63000, 70-43000, 70-53000, manufactured by Progeny, Inc..

Event Text Entries

[15607573] Certain batches of collimators built in sep-oct 1999 have incorrect (shorter) screws mounting the collimator housing to the tube mounting plate. The shorter screws over time can loosen, allowing the collimator housing to sag relative to the mounting plate. If the "looseness" of the housing was not noticed by staff, the housing could eventually separate from the mounting plate.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1423380-2001-00001
MDR Report Key314486
Report Source08
Date Received2001-01-29
Date of Report2001-01-25
Date Mfgr Received2000-11-03
Device Manufacturer Date1999-09-01
Date Added to Maude2001-02-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer Street1407 BARCLAY BLVD.
Manufacturer CityBUFFALO GROVE IL 60089
Manufacturer CountryUS
Manufacturer Postal60089
Manufacturer Phone8478503800
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Remedial ActionRC
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameLINEAR SERIES COLLIMATORS
Generic NameCOLLIMATOR
Product CodeKPW
Date Received2001-01-29
Model NumberLINEAR MC150, LINEAR II, LINEAR III, LINEAR IV
Catalog Number70-63000, 70-43000, 70-53000,
Lot Number*
ID Number*
Device AvailabilityY
Device Eval'ed by MfgrY
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key304143
ManufacturerPROGENY, INC.
Manufacturer Address1407 BARCLAY BLVD. BUFFALO GROVE IL 60089 US
Baseline Brand NameLINEAR SERIES COLLIMATORS
Baseline Generic NameCOLLIMATOR
Baseline Model NoLINEAR MC150, L
Baseline Catalog No70-63000, 70-43000, 70-53000,
Baseline ID*


Patients

Patient NumberTreatmentOutcomeDate
10 2001-01-29

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.