GOUGE F/CANCELLOUS BONE GRAFT STRAIG W/1 399.620

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,07 report with the FDA on 2013-06-04 for GOUGE F/CANCELLOUS BONE GRAFT STRAIG W/1 399.620 manufactured by Synthes Gmbh.

Event Text Entries

[3494990] It was reported that the instrument has never been used, but after sterilization at the hospital they discovered a spot at the top of instrument. This is report 1 of 1 for complaint (b)(4).
Patient Sequence No: 1, Text Type: D, B5


[10792107] Synthes is submitting this report as a result of remediation activities related to fda warning letter dated february 2012. Device(s) listed in this report is (are) used for treatment, not diagnosis. Any additional information received regarding this event after filing this report shall be filed on a supplemental mdr. A review of the device history records was performed and no complaint related issues were found. The additional evaluation revealed that there is indeed a discoloration mark on the inside of the cutting edge. Obviously the discoloration has not been spotted before sterilization. We can suppose that the tip was not dry enough and caused the discoloration/rust. Unfortunately we are not able to comprehend the mentioned problem. The manufacturing documents were reviewed and no discrepancies to the specifications were found. No product fault could be detected.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number8030965-2013-02588
MDR Report Key3145813
Report Source01,07
Date Received2013-06-04
Date of Report2011-02-22
Date Mfgr Received2011-02-22
Date Added to Maude2013-08-23
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactT. MCCARRON
Manufacturer Street1302 WRIGHTS LANE EAST
Manufacturer CityWEST CHESTER PA 19380
Manufacturer CountryUS
Manufacturer Postal19380
Manufacturer Phone8006207025
Manufacturer G1SYNTHES GMBH
Manufacturer StreetEIMATTSTRASSE 3 CH-4436
Manufacturer CityOBERDORF
Manufacturer CountrySZ
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameGOUGE F/CANCELLOUS BONE GRAFT STRAIG W/1
Product CodeGDH
Date Received2013-06-04
Catalog Number399.620
Lot Number5906287
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSYNTHES GMBH
Manufacturer AddressEIMATTSTRASSE 3 CH-4436 OBERDORF SZ


Patients

Patient NumberTreatmentOutcomeDate
10 2013-06-04

© 2026 FDA.report