MULTISTIX 8 SG 10627148

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2013-05-10 for MULTISTIX 8 SG 10627148 manufactured by Siemens Healthcare Diagnostics Inc..

Event Text Entries

[3443558] Customer reported false negative results for leukocytes on multistix 8sg strips. There was no report of injury as a result of this event.
Patient Sequence No: 1, Text Type: D, B5


[10783980] The discordant results were due to operator error. Siemens investigation found that the site was using a non-siemens instrument to read the siemens multistix 8sg strips. As stated in the multistix 8sg ifu: siemens reagent strips are ready to use upon removal from the bottle and the entire reagent strip is disposable. The strips may be read visually, requiring no additional laboratory equipment for testing. The strips can also be read instrumentally, using the clinitek family of urine chemistry analyzers and the appropriate software.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1217157-2013-00076
MDR Report Key3163073
Report Source07
Date Received2013-05-10
Date of Report2013-04-22
Date of Event2013-04-03
Date Mfgr Received2013-04-22
Date Added to Maude2013-06-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationPHARMACIST
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactSTEVEN ANDBERG
Manufacturer Street2 EDGEWATER DR.
Manufacturer CityNORWOOD MA 02062
Manufacturer CountryUS
Manufacturer Postal02062
Manufacturer Phone7812693655
Manufacturer G1KIMBELL ELECTRONICS POLAND SP.Z O.O
Manufacturer StreetUI. POZNANSKA 1/C
Manufacturer CityTARNOWO PODGORNE 62080
Manufacturer CountryPL
Manufacturer Postal Code62080
Single Use0
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameMULTISTIX 8 SG
Generic NameMULTISTIX 8 SG
Product CodeJIO
Date Received2013-05-10
Catalog Number10627148
Lot Number207056
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC.
Manufacturer Address511 BENEDICT AVE TARRYTOWN NY 10591 US 10591


Patients

Patient NumberTreatmentOutcomeDate
10 2013-05-10

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.