STERNUM BLADE 0298097100

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2013-06-18 for STERNUM BLADE 0298097100 manufactured by Stryker Instruments-kalamazoo.

Event Text Entries

[3553096] It was reported that during a surgical procedure, 2 blades broke at the mount. It was also reported that there was a minimal delay as a result of this event. It was further reported that there were no adverse consequences and the procedure was completed successfully.
Patient Sequence No: 1, Text Type: D, B5


[10855315] Manufacturing records were reviewed, no issues were identified which may have contributed to this event. The returned blades were measured where possible and all critical specifications were met. Investigation results indicate the break on the blades were consistent with the blades being bent towards the blade guard while cutting. This impact force combined with a bending force potentially caused the blades to break; however, this cannot be confirmed. The root cause is undetermined.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number0001811755-2013-01425
MDR Report Key3176013
Report Source06
Date Received2013-06-18
Date of Report2013-05-21
Date of Event2013-05-16
Date Mfgr Received2013-05-21
Device Manufacturer Date2012-09-08
Date Added to Maude2013-06-18
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR. DERVILLIA MURPHY
Manufacturer StreetINSTRUMENTS DIVISION CARRIGTWOHILL BUS. & TECH PARK
Manufacturer CityCARRIGTWOHILL NA
Manufacturer PostalNA
Manufacturer Phone214532900
Manufacturer G1STRYKER INSTRUMENTS-IRELAND
Manufacturer StreetINSTRUMENTS DIVISION CARRIGTWOHILL BUS. & TECH PARK
Manufacturer CityCARRIGTWOHILL NA
Manufacturer Postal CodeNA
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSTERNUM BLADE
Generic NameBLADE, SAW, SURGICAL, CARDIOVASCULAR
Product CodeDWH
Date Received2013-06-18
Returned To Mfg2013-06-04
Catalog Number0298097100
Lot Number12251027
Device Expiration Date2017-09-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSTRYKER INSTRUMENTS-KALAMAZOO
Manufacturer Address4100 EAST MILHAM AVENUE KALAMAZOO MI 49001 US 49001


Patients

Patient NumberTreatmentOutcomeDate
10 2013-06-18

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.