CHISEL BOX CUTTING H7 03.820.124

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,07 report with the FDA on 2013-06-20 for CHISEL BOX CUTTING H7 03.820.124 manufactured by Synthes Gmbh.

Event Text Entries

[3519214] It was reported that the chisels are rusty. This is report 2 of 2 for complaint (b)(4).
Patient Sequence No: 1, Text Type: D, B5


[10746576] Synthes is submitting this report as a result of remediation activities related to fda warning letter dated february 2012. Device(s) listed in this report is (are) used for treatment, not diagnosis. Any additional information received regarding this event after filing this report shall be filed on a supplemental mdr. A review of the device history records was performed and no complaint related issues were found. The evaluation revealed that inspection of the two chisels included in this complaint reveals that the visible traces on the surface are not only composed of rust but also include organic residues. We suspect that these deposits are due to residues left behind after the cleaning and sterilization process. In regard to rust formation, it can be generally stated that all materials, even so-called rust-free steel, is only rustproof to a certain extent. The material remains rustproof only as long as it is always dried immediately and stored in a dry place. No product defect was detected.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number8030965-2013-03760
MDR Report Key3180687
Report Source01,07
Date Received2013-06-20
Date of Report2011-06-06
Date Mfgr Received2011-06-06
Device Manufacturer Date2007-12-18
Date Added to Maude2013-09-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactT. MCCARRON
Manufacturer Street1302 WRIGHTS LANE EAST
Manufacturer CityWEST CHESTER PA 19380
Manufacturer CountryUS
Manufacturer Postal19380
Manufacturer Phone8006207025
Manufacturer G1SYNTHES GMBH
Manufacturer StreetEIMATTSTRASSE 3 CH-4436
Manufacturer CityOBERDORF
Manufacturer CountrySZ
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCHISEL BOX CUTTING H7
Product CodeEML
Date Received2013-06-20
Returned To Mfg2011-06-10
Catalog Number03.820.124
Lot NumberA7QA50
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSYNTHES GMBH
Manufacturer AddressEIMATTSTRASSE 3 CH-4436 OBERDORF SZ


Patients

Patient NumberTreatmentOutcomeDate
10 2013-06-20