CHISEL-FLAT-STRAIGHT W/8 397.970

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,07 report with the FDA on 2013-06-20 for CHISEL-FLAT-STRAIGHT W/8 397.970 manufactured by Synthes Gmbh.

Event Text Entries

[3518254] It was reported that the chisel broke during a total knee replacement surgery. The chisel was being used on periprosthetic cement. Product only used less than one year and was broken. Customer complained that the product is not durable. This is report 1 of 1 for complaint (b)(4).
Patient Sequence No: 1, Text Type: D, B5


[10747331] Additional narrative: synthes is submitting this report as a result of remediation activities related to fda warning letter dated february 2012. Device(s) listed in this report is (are) used for treatment, not diagnosis. Any additional information received regarding this event after filing this report shall be filed on a supplemental mdr. A review of the device history records was performed and no complaint related issues were found. The evaluation revealed that the blade of the chisel was partially broken off. All articles have been distributed worldwide and we are not aware off any other similar breakages related to the mentioned lot number. The microscopic view of the broken surfaces does not show any anomalies of material structure. The measurable dimensions were checked and found to be in compliance with the technical drawings and ao asif specification. The instrument meets fully to our specifications. No product fault could be detected.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number8030965-2013-03694
MDR Report Key3180699
Report Source01,07
Date Received2013-06-20
Date of Report2011-08-03
Date Mfgr Received2011-08-03
Device Manufacturer Date2011-05-16
Date Added to Maude2013-09-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactT. MCCARRON
Manufacturer Street1302 WRIGHTS LANE EAST
Manufacturer CityWEST CHESTER PA 19380
Manufacturer CountryUS
Manufacturer Postal19380
Manufacturer Phone8006207025
Manufacturer G1SYNTHES GMBH
Manufacturer StreetEIMATTSTRASSE 3 CH-4436
Manufacturer CityOBERDORF
Manufacturer CountrySZ
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCHISEL-FLAT-STRAIGHT W/8
Product CodeEML
Date Received2013-06-20
Catalog Number397.970
Lot Number2715973
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSYNTHES GMBH
Manufacturer AddressEIMATTSTRASSE 3 CH-4436 OBERDORF SZ


Patients

Patient NumberTreatmentOutcomeDate
10 2013-06-20