DIMENSION? XPAND PLUS WITH HM

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2013-06-21 for DIMENSION? XPAND PLUS WITH HM manufactured by Siemens Healthcare Diagnostics Inc.

Event Text Entries

[3468830] Discordant falsely elevated hemoglobin a1c (hb1c) patient results were obtained on three patients on a dimension xpand plus with hm system. The results were reported to the physician(s) who did not question the elevated results. After obtaining an acceptable calibration with lot ga4028, the customer re-tested the patient samples and the corrected results were reported. There are no known reports of patient intervention or adverse health consequences due to the discordant high hemoglobin a1c (hb1c) patient results.
Patient Sequence No: 1, Text Type: D, B5


[10862324] Siemens was contacted by the customer who requested assistance with calibrating hemoglobin a1c (hb1c). The customer received arithmetic errors on the raw calibration data for lot ga4112. The customer calculated the data and accepted a bad calibration, therefore the tsc determined that the cause of the discordant hb1c results was user error. The falsely elevated hb1c results were caused by testing samples with a bad calibration. The issue was resolved when the customer re-calibrated using an alternate flex lot. The instrument is performing within specifications. Further evaluation of the device is not required.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1226181-2013-00280
MDR Report Key3183160
Report Source05,06
Date Received2013-06-21
Date of Report2013-05-29
Date of Event2013-05-29
Date Mfgr Received2013-05-29
Device Manufacturer Date2010-03-04
Date Added to Maude2013-09-13
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactJOHN NELSON
Manufacturer StreetSIEMENS HEALTHCARE DIAGNOSTICS 511 BENEDICT AVENUE
Manufacturer CityTARRYTOWN NY 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145242530
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS INC
Manufacturer Street101 SILVERMINE ROAD
Manufacturer CityBROOKFIELD CT 06804
Manufacturer CountryUS
Manufacturer Postal Code06804
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDIMENSION? XPAND PLUS WITH HM
Generic NameIMMUNOASSAY ANALYZER
Product CodePDJ
Date Received2013-06-21
Model NumberDIMENSION? XPAND PLUS WITH HM
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Age3 YR
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC
Manufacturer Address101 SILVERMINE ROAD BROOKFIELD CT 06804 US 06804


Patients

Patient NumberTreatmentOutcomeDate
10 2013-06-21

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.