THINPREP IMAGING SYSTEM

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05 report with the FDA on 2013-06-20 for THINPREP IMAGING SYSTEM manufactured by Hologic Inc.

Event Text Entries

[3515427] Lab personnel reported a case with observed abnormal cells that were not located in the 22 fields of view (fov) selected by the imager. Cytology application specialist (cas) reviewed slides on the review scope as an unknown mixed with other slides. Two cases presented had no triggers in the 22 fobs to prompt an autoscan. Full review of slide during qc showed abnormal cells outside the fovs. Cas reviewed this case and discussed with lap personnel. Cas agreed there were no changes seen in 22 fov to prompt an autoscan. Please comment on number of abnormal cells present: rare. Please comment on the overall appearance of the slide: (b)(6): the background of the slide was cytolysis there was one group of lsil; (b)(6): the background was clean with superficial and intermediate cells. There was one large keratinized cell with a large abnormal nucleus. The nucleus looked a little degenerate. Customer did not request slide to be reviewed at corporate. Customer will continue to hold questionable cases for cas review. Customer did not require follow up from customer support services.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1222780-2013-00125
MDR Report Key3187970
Report Source05
Date Received2013-06-20
Date of Report2013-05-21
Date of Event2013-05-21
Date Mfgr Received2013-05-21
Date Added to Maude2013-06-27
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactRANDALL COVILL, MGR
Manufacturer Street250 CAMPUS DRIVE
Manufacturer CityMARLBOROUGH MA 01752
Manufacturer CountryUS
Manufacturer Postal01752
Manufacturer Phone5082638853
Single Use0
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTHINPREP IMAGING SYSTEM
Generic NameAUTOMATED MICROSCOPE FOR CYTOLOGY
Product CodeMNM
Date Received2013-06-20
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerHOLOGIC INC
Manufacturer AddressMARLBOROUGH MA 01752 US 01752


Patients

Patient NumberTreatmentOutcomeDate
101. Life Threatening 2013-06-20

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.