DIMENSION RXL MAX W/O HM DIMENSION RXL MAX W/ HM

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2013-06-28 for DIMENSION RXL MAX W/O HM DIMENSION RXL MAX W/ HM manufactured by Siemens Healthcare Diagnostics Inc..

Event Text Entries

[15205833] A discordant, falsely elevated gentamicin result was obtained for one patient sample on a dimension rxl max w/o hm instrument. The discordant result was reported to the physician(s), who questioned the result. The sample was then rerun twice on this instrument and each time resulted lower. The rerun results were reported to the physician(s). There are no known reports of patient intervention or adverse health consequences due to the discordant, falsely elevated gentamicin result.
Patient Sequence No: 1, Text Type: D, B5


[15541293] The customer contacted the siemens technical solutions center (tsc). After evaluation of the instrument data, the tsc specialist discovered that the r2 probe needed replacement. The customer replaced the r2 probe and the instrument is functioning as expected. The cause of the discordant, falsely elevated gentamicin result is due to a malfunction of the r2 probe. This instrument is performing according to specifications. No further evaluation of this device is required.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1226181-2013-00287
MDR Report Key3195725
Report Source05,06
Date Received2013-06-28
Date of Report2013-06-03
Date of Event2013-06-03
Date Mfgr Received2013-06-03
Device Manufacturer Date2004-02-24
Date Added to Maude2013-09-20
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactCASSANDRA KOCSIS
Manufacturer Street511 BENEDICT AVENUE
Manufacturer CityTARRYTOWN NY 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145242687
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS INC.
Manufacturer Street101 SILVERMINE ROAD
Manufacturer CityBROOKFIELD CT 06804
Manufacturer CountryUS
Manufacturer Postal Code06804
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDIMENSION RXL MAX W/O HM
Generic NameCLINICAL CHEMISTRY ANALYZER
Product CodeLCD
Date Received2013-06-28
Model NumberDIMENSION RXL MAX W/ HM
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC.
Manufacturer Address101 SILVERMINE ROAD BROOKFIELD CT 06804 US 06804


Patients

Patient NumberTreatmentOutcomeDate
10 2013-06-28

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.