DEVICE 8252200 NIM-RESPONSE 2.0 INTERFAC

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 03,06 report with the FDA on 2013-06-28 for DEVICE 8252200 NIM-RESPONSE 2.0 INTERFAC manufactured by Medtronic Xomed, Inc..

Event Text Entries

[21293183] It was reported the box has to be situated in the right position in order to work; works intermittently.
Patient Sequence No: 1, Text Type: D, B5


[21483320] This device is used for therapeutic purposes. (b)(4). The returned device was evaluated by quality engineer. The mainframe was powered on with the interface, simulator, and probe c onnected. A monitoring mode using all available channels was selected. Each channel was stimulated in turn using stim 1 and stim 2; proper responses and measured currents were seen on any channels with each stim circuit. However, a muting warning intermittently appeared. The interface was unplugged and then plugged back in. The electrode check did not pick up any impedance readings.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1045254-2013-00436
MDR Report Key3195741
Report Source03,06
Date Received2013-06-28
Date of Report2013-06-03
Date Mfgr Received2013-06-03
Device Manufacturer Date2005-07-14
Date Added to Maude2013-08-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactCHRISTY CAIN
Manufacturer Street6743 SOUTHPOINT DRIVE NORTH
Manufacturer CityJACKSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal32216
Manufacturer Phone9043328353
Manufacturer G1MEDTRONIC XOMED, INC.
Manufacturer Street6743 SOUTHPOINT DRIVE NORTH
Manufacturer CityJACKSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal Code32216
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDEVICE 8252200 NIM-RESPONSE 2.0 INTERFAC
Generic NameELECTROMYOGRAPH, DIAGNOSTIC
Product CodeIKN
Date Received2013-06-28
Returned To Mfg2013-06-07
Model Number8252200
Catalog Number8252200
Lot Number39290600
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC XOMED, INC.