MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,07 report with the FDA on 2013-06-25 for TC BABY-CRILE-WOOD NDL HLDR SERR 150MM BM013R manufactured by Aesculap Ag & Co Kg.
[17928798]
Country of complaint: (b)(6). The tip of the needle holder broke off intraoperatively. The possibility, that the tip fell into the pt cannot be ruled out.
Patient Sequence No: 1, Text Type: D, B5
[18153767]
Evaluation: needle holders which have been repaired, show a little bigger tendency to break than brand new ones, if the carbide metal was changed. This problem is well known to us. But nevertheless even the strength of a repaired needle holder is strong enough to last with proper handling. Whether the handling was improper or not cannot be ascertained. It concerns a forced fracture caused by a mechanical overload situation. Despite several requests we did not receive info regarding the pt's condition.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 2916714-2013-00040 |
MDR Report Key | 3196805 |
Report Source | 01,07 |
Date Received | 2013-06-25 |
Date of Report | 2013-05-22 |
Date Mfgr Received | 2013-03-21 |
Date Added to Maude | 2013-09-18 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 0 |
Initial Report to FDA | 3 |
Report to FDA | 0 |
Event Location | 0 |
Manufacturer Contact | MICHELLE LINK |
Manufacturer Street | 615 LAMBERT POINTE DR |
Manufacturer City | HAZELWOOD MO 63042 |
Manufacturer Country | US |
Manufacturer Postal | 63042 |
Manufacturer Phone | 3145515938 |
Manufacturer G1 | AESCULAP AG & CO. KG |
Manufacturer Street | P. O. BOX 40 |
Manufacturer City | TUTTLINGEN |
Manufacturer Country | GM |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | TC BABY-CRILE-WOOD NDL HLDR SERR 150MM |
Product Code | HXK |
Date Received | 2013-06-25 |
Returned To Mfg | 2013-03-27 |
Model Number | BM013R |
Catalog Number | BM013R |
Operator | HEALTH PROFESSIONAL |
Device Availability | Y |
Device Age | DA |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | AESCULAP AG & CO KG |
Manufacturer Address | TUTTLINGEN 78532 78532 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Other | 2013-06-25 |