SMART STAPES PISTON 70145923

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2013-06-27 for SMART STAPES PISTON 70145923 manufactured by Gyrus Acmi, Inc..

Event Text Entries

[3631954] It was reported that during a surgical procedure the smart stapes piston was implanted in the pts ear. The surgeon noticed the piston was falling backwards against the facial nerve when placed initially and decided to remove it. During the removal procedure, the piston separated and fell into the pts middle ear. Most of the foot plate was taken out. The surgeon stated that it was difficult to retrieve the piston as it was against the facial nerve and had to remove 2mm of the vestibule. The incident resulted in extended surgical time as well as leaving the pt with some hearing loss and experience of dizziness since the procedure. The intended procedure was completed. Additional treatment was required, the graft from the pt's hand was used to replaced the 2mm vestibule. A new piston was implanted in the pts ear.
Patient Sequence No: 1, Text Type: D, B5


[10971747] An investigation was performed on the returned product. The investigation revealed that the product was not manufactured to specifications. A review of the dhr and previous complaint history has been done. Cyrus acmi will continue to monitor future reports.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1037007-2013-00010
MDR Report Key3199452
Report Source07
Date Received2013-06-27
Date of Report2013-05-30
Date of Event2013-05-13
Date Mfgr Received2013-05-30
Date Added to Maude2013-07-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactTERRENCE SULLIVAN
Manufacturer Street136 TURNPIKE ROAD
Manufacturer CitySOUTHBOROUGH MA 01772
Manufacturer CountryUS
Manufacturer Postal01772
Manufacturer Phone5088042739
Manufacturer G1GYRUS ACMI, INC.
Manufacturer Street2925 APPLING RD.
Manufacturer CityBARTLETT TN 38133390
Manufacturer CountryUS
Manufacturer Postal Code38133 3901
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameSMART STAPES PISTON
Generic NameSMART STAPES PISTON
Product CodeETB
Date Received2013-06-27
Returned To Mfg2013-06-14
Model Number70145923
Catalog Number70145923
Lot NumberMH51432
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerGYRUS ACMI, INC.
Manufacturer AddressBARTLETT TN US


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization; 2. Required No Informationntervention 2013-06-27