AFP * 857000001

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2001-03-12 for AFP * 857000001 manufactured by Afp Imaging Corp.

Event Text Entries

[197349] Soft film cassettes for panoramic x-ray systems are not light tight. X-ray film is fogged within 5 mins when casette is removed from the darkroom, complainant has worked with supplier and has determined that 10 units all had the same problem. At the request of afp, a comparison was made with a competitor's (gendex) product; the gendex product was light tight. Afp cassette is used with lanex regular screens and e-mat-g film. The 6/12" size is used on a kaykor panaramic x-ray system.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report NumberMW4002959
MDR Report Key320254
Date Received2001-03-12
Date of Report2001-02-15
Date Added to Maude2001-03-15
Event Key0
Report Source CodeVoluntary report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA0
Report to FDA0
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameAFP
Generic NameSOFT FILM CASSETTE, 6X12" FOR PANORAMIC X-RAY SYSTEM
Product CodeIXA
Date Received2001-03-12
Model Number*
Catalog Number857000001
Lot Number*
ID Number*
Device Availability*
Implant FlagN
Date Removed*
Device Sequence No1
Device Event Key309778
ManufacturerAFP IMAGING CORP
Manufacturer Address250 CLEARBROOK RD ELMSFORD NY 10523 US


Patients

Patient NumberTreatmentOutcomeDate
10 2001-03-12

© 2026 FDA.report