MAUDE MDR 3209568

MDR report key
3209568
Report number
3007056120-2013-00009
Event key
0
Event type
3
Date received
2013-07-02
Adverse event
3
Product problem
3
Patients in event
0
Reporter occupation
0
Health professional
3
Initial report to FDA
3
Event location
0

Manufacturer Contact#

Contact
CHRIS FERGUSON
Address
191 WYNGATE DR MONROEVILLE PA 15146 US
Report source
M
Manufacturer link flag
Y

Devices#

Seq, Brand, Generic table
SeqBrandGenericManufacturerProduct codeModelCatalogLotPMA510(k)ImplantEvaluatedAvailability
1SMARTMONITOR 2APNEA MONITORPHILIPS RESPIRONICS - CHMVFLS4002N N

Patients#

Sequence, Received, Treatment table
SequenceReceivedTreatmentOutcome
12013-07-020

Event Narratives#

D

Patient 1

CHILDREN'S MEDICAL VENTURES (CHMV) RECEIVED A COMPLAINT REPORT FROM A DURABLE MEDICAL EQUIPMENT (DME) SUPPLIER STATING THAT A SMARTMONITOR DEVICE WAS FOUND TO HAVE NO AUDIBLE ALARM DURING CHECK-OUT PROCEDURAL TESTING. THE DEVICE WAS REPORTEDLY NOT IN PT USE AT THE TIME, THERE WAS NO DEATH OR SERIOUS INJURY ASSOCIATED WITH THE ALLEGED COMPLAINT CONDITION.

N

Patient 1

(B)(4). THE MFR RECEIVED THE DEVICE FOR EVAL AND THE CUSTOMER COMPLAINT WAS CONFIRMED. THE DEVICE WAS FOUND TO HAVE NO AUDIBLE ALAR. HOWEVER, THE DESIGN INCORPORATED LED VISUAL INDICATOR FUNCTIONED TO SPEC. THE UNIT ALARMED TO SPEC WITH THE TEST ALARM MODULE. THE CAUSE OF THE FAILURE WAS ISOLATED TO THE ALARM MODULE. THE ALARM MODULE WAS FORWARDED TO SUPPLIER QUALITY ENGINEERING FOR FURTHER INVESTIGATION. SMARTMONITOR 2 IS DESIGNED TO MONITOR AND RECORD PT'S BREATHING (RESPIRATION), HEART (CARDIAC) ACTIVITY. THE MONITOR ALERTS YOU IF ANY OF THESE ACTIVITIES EXCEEDS AND/OR HEART ACTIVITY ARE NOT WITHIN THESE SET BOUNDARIES, AN INDICATOR LIGHT COMES ON AND AN ALARM SOUNDS. A SUPPLIER CORRECTIVE ACTION HAS BEEN INITIATED WITH THE MFR OF THE ALARM MODULE TO INVESTIGATE THE RECORDED FAILURE. ONCE THE INVESTIGATION HAS BEEN COMPLETED, A F/U REPORT WILL BE FILED ON DETAIL THE CONCLUSIONS REACHED DURING THE INVESTIGATION.