CONSOLE COMMAND SYSTEM NONE 5D-296-10-40K

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 1992-07-09 for CONSOLE COMMAND SYSTEM NONE 5D-296-10-40K manufactured by Stryker Instruments.

Event Text Entries

[2196] In the process of doing oral surgery, thge distal portion of the handpiece (burr holder) became hot and burned patient. Injury assessed by a physician consultant and burn treatment starteddevice not labeled for single use. Patient medical status prior to event: satisfactory condition. There was not multiple patient involvement. Device serviced in accordance with service schedule. Date last serviced: 01-aug-91. Service provided by: user facility biomedical/bioengineering department. Service records available. No imminent hazard to public health claimed. Device used as labeled/intended. Device was evaluated after the event. Method of evaluation: actual device involved in incident was evaluated, performance tests performed. Results of evaluation: component failure, telemetry failure, none or unknown, component failure. Conclusion: device failure directly caused event. Certainty of device as cause of or contributor to event: yes. Corrective actions: device permanently removed from service. The device was not destroyed/disposed of.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3211
MDR Report Key3211
Date Received1992-07-09
Date of Report1992-03-23
Date of Event1992-03-17
Date Facility Aware1992-03-17
Report Date1992-03-23
Date Reported to Mfgr1992-03-17
Date Added to Maude1993-04-12
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional0
Initial Report to FDA0
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameCONSOLE COMMAND SYSTEM
Generic NameSURGICAL DRILL
Product CodeHBD
Date Received1992-07-09
Model NumberNONE
Catalog Number5D-296-10-40K
Lot NumberNONE
ID Number28864
OperatorOTHER HEALTH CARE PROFESSIONAL
Device AvailabilityY
Device Age01-JUL-91
Implant FlagN
Device Sequence No1
Device Event Key2995
ManufacturerSTRYKER INSTRUMENTS


Patients

Patient NumberTreatmentOutcomeDate
101. Other 1992-07-09