EAGLE EO MONITOR

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2013-07-08 for EAGLE EO MONITOR manufactured by Steris Mexico, S. De R.l. De C.v..

Event Text Entries

[21108159] The user facility reported that their eagle eo monitor was alarming? High leak rate of 9. 12?. The surgical department was evacuated and emergency services were contacted. No injuries to hospital staff or patients were reported. No procedural delays or cancellations occurred.
Patient Sequence No: 1, Text Type: D, B5


[21158883] A steris service technician arrived at the facility to inspect the unit. The technician confirmed that no eto sterilizers were running on the day of the reported event. The monitor was found to be operating to specification. Emergency services were contacted by the user facility. The eo alarm turned off before they arrived. The monitor alarmed for one full cycle only. The technician confirmed the alarm initiated by a non-eo odor as the monitor is located in area with multiple devices and frequent traffic.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3005899764-2013-00067
MDR Report Key3211373
Report Source06
Date Received2013-07-08
Date of Report2013-07-08
Date of Event2013-06-09
Date Mfgr Received2013-06-09
Date Added to Maude2013-07-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS KATHRYN CADORETTE
Manufacturer Street5960 HEISLEY ROAD
Manufacturer CityMENTOR OH 44060
Manufacturer CountryUS
Manufacturer Postal44060
Manufacturer Phone4403927231
Manufacturer G1STERIS MEXICO, S. DE R.L. DE C.V.
Manufacturer StreetAVENIDA AVANTE 790 PARQUE INDUSTRIAL GUADALUPE
Manufacturer CityGUADALUPE, NUEVO LEON 67190
Manufacturer CountryMX
Manufacturer Postal Code67190
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameEAGLE EO MONITOR
Generic NameSTERILIZER MONITOR
Product CodeFLF
Date Received2013-07-08
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSTERIS MEXICO, S. DE R.L. DE C.V.
Manufacturer AddressAVENIDA AVANTE 790 PARQUE INDUSTRIAL GUADALUPE GUADALUPE, NUEVO LEON 67190 MX 67190


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2013-07-08

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.