ADVIA 120 067-A004

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05,06 report with the FDA on 2013-07-12 for ADVIA 120 067-A004 manufactured by Siemens Healthcare Diagnostics Inc..

Event Text Entries

[3652169] The customer reported that hemoglobin reagent leaked out of a reagent pump on the advia 120 instrument, causing an electrical short that affected power to the customer lab. There were no injuries reported, no known reports of medical intervention or adverse health consequences and no known delay in patient testing due to the leaking reagent pump.
Patient Sequence No: 1, Text Type: D, B5


[10997044] A siemens healthcare diagnostics field service engineer (fse) was dispatched to the customer site for instrument evaluation. After evaluating the instrument, the fse determined that the cause for the hemoglobin reagent leak was due to a malfunction in the reagent pump membrane. The fse replaced the reagent pump and electrical components affected by the leak. The analyzer is operating according to specifications. No further evaluation of the analyzer is required.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2432235-2013-00320
MDR Report Key3222877
Report Source01,05,06
Date Received2013-07-12
Date of Report2013-06-19
Date of Event2013-06-13
Date Mfgr Received2013-06-19
Device Manufacturer Date2003-03-22
Date Added to Maude2013-10-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactJOHN NELSON
Manufacturer Street511 BENEDICT AVENUE
Manufacturer CityTARRYTOWN NY 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145242530
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING, LTD.
Manufacturer StreetCHAPEL LANE SWORDS
Manufacturer CityDUBLIN,
Manufacturer CountryEI
Single Use3
Remedial ActionRP
Previous Use Code3
Removal Correction NumberN/A
Event Type3
Type of Report3

Device Details

Brand NameADVIA 120
Generic NameHEMATOLOGY ANALYZER
Product CodeGKL
Date Received2013-07-12
Model NumberADVIA 120
Catalog Number067-A004
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Age10 YR
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC.
Manufacturer Address511 BENEDICT AVENUE TARRYTOWN NY 10591 US 10591


Patients

Patient NumberTreatmentOutcomeDate
10 2013-07-12

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.