TREROTOLA GASTROSTOMY TRACT DILATION SET GTDS-24-8-10-0

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,07 report with the FDA on 2001-03-28 for TREROTOLA GASTROSTOMY TRACT DILATION SET GTDS-24-8-10-0 manufactured by Cook, Inc..

Event Text Entries

[224555] With an amplatz stiff wire and the balloon catheter fixed securely, the staff was attempting to place the sheath over the inflated balloon catheter and wire combination. During the attempt, the catheter broke at the junction of the balloon. The balloon remained partially inflated and could not be pulled back through the sheath. The broken catheter segment was snared and retrieved without further difficulty.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1820334-2001-00028
MDR Report Key322641
Report Source05,07
Date Received2001-03-28
Date of Report2001-02-28
Report Date2001-02-28
Date Mfgr Received2001-02-28
Device Manufacturer Date2000-12-01
Date Added to Maude2001-03-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer ContactDENNIS TROY
Manufacturer Street925 SOUTH CURRY PIKE
Manufacturer CityBLOOMINGTON IN 47403
Manufacturer CountryUS
Manufacturer Postal47403
Manufacturer Phone8123392235
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTREROTOLA GASTROSTOMY TRACT DILATION SET
Generic NameSPECIAL SET
Product CodeGCC
Date Received2001-03-28
Model NumberNA
Catalog NumberGTDS-24-8-10-0
Lot NumberF989039
ID NumberNA
Device Expiration Date2003-11-01
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Age2 MO
Device Eval'ed by MfgrY
Implant FlagN
Date RemovedA
Device Sequence No1
Device Event Key312104
ManufacturerCOOK, INC.
Manufacturer Address925 SOUTH CURRY PIKE BLOOMINGTON IN 47403 US
Baseline Brand NameTREROTOLA GASTROSTOMY TRACT DILATION SET
Baseline Generic NameSPECIAL SET
Baseline Model NoNA
Baseline Catalog NoGTDS-24-8-10-0
Baseline IDNA


Patients

Patient NumberTreatmentOutcomeDate
101. Required No Informationntervention 2001-03-28