FREEDOM EVO-2 200 10641200

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2013-07-15 for FREEDOM EVO-2 200 10641200 manufactured by Tecan Schweiz Ag.

Event Text Entries

[3654458] An evo instrument operator reached with their hand into the instrument while in operation and the pick and place arm (pnp gripper) hit their hand pinning it to the inside of the safety shield resulting in a bruise. On the day of the event, the operator's hand was swollen and hurting. There were no cuts or broken skin and the operator did not seek first aid or medical treatment. The operator was a contract employee of the laboratory and informed laboratory staff 3 days later of the event. The laboratory reported to tecan the event on (b)(6). At that time, the operator indicated the swelling had extended to the arm. Lab staff advised operator to consult a physician. No further information has become available about the nature or extent of the injuries or treatment required. Tecan made multiple attempts to obtain this information. The operator's injury was the result of disabled door locks which allowed the operator to reach inside a running instrument.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1037985-2013-00001
MDR Report Key3231315
Date Received2013-07-15
Date of Report2013-07-15
Date of Event2013-06-24
Date Facility Aware2013-06-24
Report Date2013-07-15
Date Reported to FDA2013-07-15
Date Reported to Mfgr2013-06-25
Date Added to Maude2013-07-18
Event Key0
Report Source CodeDistributor report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Manufacturer Street4022 STIRRUP CREEK RD.
Manufacturer CityRESEARCH TRIANGLE PARK NC 27709
Manufacturer CountryUS
Manufacturer Postal27709
Manufacturer G1TECAN U.S.
Manufacturer Street4022 STIRRUP CREEK RD.
Manufacturer CityRESEARCH TRIANGLE PARK NC 27709
Manufacturer CountryUS
Manufacturer Postal Code27709
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameFREEDOM EVO-2 200
Generic NamePIPETTING STATION FOR CLINICAL USE
Product CodeJQW
Date Received2013-07-15
Model NumberEVO-2 200
Catalog Number10641200
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Age18 MO
Device Sequence No1
Device Event Key0
ManufacturerTECAN SCHWEIZ AG
Manufacturer AddressMANNENDORF, ZURICH SZ


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2013-07-15

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.