PROGENY VANTAGE VANTAGE WITH CEPH EXTENSION V5000C

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2013-07-18 for PROGENY VANTAGE VANTAGE WITH CEPH EXTENSION V5000C manufactured by Midmark Corporation.

Event Text Entries

[3687443] A radiologic technologist from a dental imaging center reported to a midmark sales representative that during the process of positioning a patient, a vantage panoramic x-ray unit, serial number (b)(4), would continue moving down after the "down" button on the patient table was released. The radiologic technologist removed the patient from the machine. There was only one occurrence and the machine performed normally afterwards. The midmark sales representative spoke to the radiologic technologist and confirmed there was no injury.
Patient Sequence No: 1, Text Type: D, B5


[10893305] The unit is equipped with an emergency stop switch. The user manual explains that "depressing the button will immediately halt all motor movement". A midmark sales representative performed an on-site investigation and confirmed that the vertical movement switches allowed the column to move up, down and stop upon command. The emergency stop switch was confirmed to function as intended. The vertical movement switch and the associated cable were replaced and returned to midmark for further analysis. The alleged failure could not be reproduced.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1423380-2013-00003
MDR Report Key3234347
Report Source07
Date Received2013-07-18
Date of Report2013-07-16
Date of Event2013-06-19
Date Mfgr Received2013-06-19
Device Manufacturer Date2013-05-21
Date Added to Maude2013-10-16
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRADIOLOGIC TECHNOLOGIST
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR. MIKE COYLE
Manufacturer Street675 HEATHROW DR.
Manufacturer CityLINCOLNSHIRE 60069
Manufacturer CountryUS
Manufacturer Postal60069
Manufacturer Phone8474159800
Manufacturer G1MIDMARK CORPORATION
Manufacturer Street675 HEATHROW DR.
Manufacturer CityLINCOLNSHIRE IL 60069
Manufacturer CountryUS
Manufacturer Postal Code60069
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePROGENY VANTAGE
Generic NameSYSTEM,X-RAY,EXTRAORAL SOURCE,DIGITAL
Product CodeMUH
Date Received2013-07-18
Returned To Mfg2013-07-11
Model NumberVANTAGE WITH CEPH EXTENSION
Catalog NumberV5000C
OperatorRADIOLOGIC TECHNOLOGIST
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMIDMARK CORPORATION
Manufacturer Address675 HEATHROW DR. LINCOLNSHIRE IL 60069 US 60069


Patients

Patient NumberTreatmentOutcomeDate
10 2013-07-18