GYNECARE THERMACHOICE TC003 *

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2013-04-26 for GYNECARE THERMACHOICE TC003 * manufactured by Ethicon, Inc..

Event Text Entries

[3666798] "thermachoice" machine not functioning, unable to maintain pressure. A second hand piece opened and handed off to the physician. The second hand piece was not functioning. The physician was unable to perform "thermachoice" ablation; however, elected to perform uterine ablation utilizing an alternative uterine ablation. Patient tolerated procedure well. "thermachoice" generator removed from service, tagged, biomed was notified. Hand pieces and packaging saved. Biomed checked generator/hand pieces. Generator given all clear and placed back in service. According to biomed hand pieces one had a pin hole and one seal cracked. What was the original intended procedure? Uterine ablation. Device #1is this a laboratory device or laboratory test? No.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3242464
MDR Report Key3242464
Date Received2013-04-26
Date of Report2013-04-26
Date of Event2013-04-16
Report Date2013-04-26
Date Reported to FDA2013-04-26
Date Reported to Mfgr2013-07-23
Date Added to Maude2013-07-23
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameGYNECARE THERMACHOICE
Generic NameDEVICE, THERMAL ABLATION
Product CodeMKN
Date Received2013-04-26
Model NumberTC003
Catalog Number*
Lot NumberDPMG05
ID Number*
OperatorPHYSICIAN
Device AvailabilityY
Device Age1 DY
Device Sequence No1
Device Event Key0
ManufacturerETHICON, INC.
Manufacturer AddressROUT 22 WEST SUMMERVILLE NJ 08876 US 08876


Patients

Patient NumberTreatmentOutcomeDate
10 2013-04-26

© 2026 FDA.report