TOBII C-12

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 99 report with the FDA on 2013-06-03 for TOBII C-12 manufactured by Tobii Electronics Technology Suzhou Co., Ltd..

Event Text Entries

[3661094] Wheelchair power adapter shorted out and melted. Fuse blew but there still appeared to be additional damage to wire. I asked (b)(6) to call in to tech support about it and to send us photos and/or the item for analysis. C12 does not appear to be damaged, no injuries, just concern about safety of this item.
Patient Sequence No: 1, Text Type: D, B5


[10983884] A unit was reported faulty in the field and returned to us - upon closer inspection, it seems as if a short circuit occurred. The reason is likely due to improper third party installation, resulting in a grounding fault between the wheelchair chassis and the wheelchair battery. Under this circumstance, and in the case of a short circuit, there is a risk of a heat rush which could cause a fire. We feel that the consequences could have been significant enough that we have made the decision to implement additional safety measures to the design, and retroactively fit all units in the field at no charge to the customer.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3007873415-2013-00002
MDR Report Key3253582
Report Source99
Date Received2013-06-03
Date of Report2013-05-14
Date of Event2012-01-26
Date Mfgr Received2012-01-27
Device Manufacturer Date2011-02-01
Date Added to Maude2013-08-22
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactLYLE SHERWOOD
Manufacturer Street333 ELM STREET
Manufacturer CityDEDHAM MA 02026
Manufacturer CountryUS
Manufacturer Postal02026
Single Use3
Remedial ActionMA
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTOBII
Generic NameC-12
Product CodeILQ
Date Received2013-06-03
Model NumberC-12
OperatorLAY USER/PATIENT
Device AvailabilityN
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerTOBII ELECTRONICS TECHNOLOGY SUZHOU CO., LTD.
Manufacturer AddressNO. 678, FENGTING AVE. LAND INDUSTRIAL PARK