IMMAGE? 800 IMMUNOCHEMISTRY SYSTEM A15445

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2013-07-30 for IMMAGE? 800 IMMUNOCHEMISTRY SYSTEM A15445 manufactured by Beckman Coulter.

Event Text Entries

[16551259] The customer reported discovering a leak under the sample wheel of the immage 800 immunochemistry system. The customer could not identify the source of the leak and had requested for service to evaluate the instrument. The customer stated that approximately 10 ml of fluid leaked and was contained within the instrument. The customer was wearing personal protective equipment consisting of a laboratory coat, gloves, and eye protection at the time of the event and no injury or exposure was reported. No erroneous patient results were generated and there was no change or affect to patient treatment in connection with this event.
Patient Sequence No: 1, Text Type: D, B5


[16846435] A beckman coulter field service engineer (fse) was dispatched to the customer's site. The fse discovered a leak at the t fitting which supplies wash solution to sample and reagent syringe valve. The fse proceeded to re-tighten the fitting to resolve the leak and verified repair per established procedures. Failure mode of the leak is attributed to a loose t fitting. Results: t fitting. Beckman coulter internal identifier for this report is (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number2050012-2013-00530
MDR Report Key3255435
Report Source05,06
Date Received2013-07-30
Date of Report2013-07-11
Date of Event2013-07-11
Date Mfgr Received2013-07-11
Device Manufacturer Date2013-04-09
Date Added to Maude2013-10-24
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS. DUNG NGUYEN
Manufacturer Street250 S. KRAEMER BOULEVARD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal92821
Manufacturer Phone7149614941
Manufacturer G1BECKMAN COULTER
Manufacturer Street250 S. KRAEMER BLVD
Manufacturer CityBREA CA 92821
Manufacturer CountryUS
Manufacturer Postal Code92821
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameIMMAGE? 800 IMMUNOCHEMISTRY SYSTEM
Generic NameNEPHELOMETER, FOR CLINICAL USE
Product CodeJQX
Date Received2013-07-30
Model NumberNA
Catalog NumberA15445
Lot NumberNA
ID NumberSW VERSION 1.6
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerBECKMAN COULTER
Manufacturer Address250 S. KRAEMER BLVD BREA CA 92821 US 92821


Patients

Patient NumberTreatmentOutcomeDate
10 2013-07-30

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.