GETINGE 46-SERIES 46-4

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 99 report with the FDA on 2013-07-17 for GETINGE 46-SERIES 46-4 manufactured by Getinge Disinfection Ab.

Event Text Entries

[15202797] The fire alarms went off at 15:30 on (b)(6) 2010. The hospital used the washer last approx 21:00 (b)(6) 2010. No other detail of the report is available at this time.
Patient Sequence No: 1, Text Type: D, B5


[15787728] Due to an incident resulting in mdr #9616031-2013-00001, a retrospective review of similar complaints identified this incident having occurred with no mdr submitted. The device investigation concluded that the heating element was activated without water in the chamber and that the overheat protection failed. A correction of affected 46-4 washer-disinfectors was initiated on (b)(6) 2013 and report to fda in preparation. The report of voluntary correction will be submitted on the agency by or prior to (b)(6) 2013.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9616031-2013-00002
MDR Report Key3264606
Report Source99
Date Received2013-07-17
Date of Report2013-07-17
Date of Event2010-08-14
Date Mfgr Received2010-08-14
Device Manufacturer Date2010-08-25
Date Added to Maude2013-10-08
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactANN WHEELER(US AGENT)
Manufacturer Street1777 E HENRIETTA RD
Manufacturer CityROCHESTER NY 14623
Manufacturer CountryUS
Manufacturer Postal14623
Manufacturer Phone5852725036
Manufacturer G1GETINGE DISINFECTION AB
Manufacturer StreetLJUNGADALSGATAN 11
Manufacturer CityVAXJO 35115
Manufacturer CountrySW
Manufacturer Postal Code35115
Single Use3
Remedial ActionRC
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameGETINGE 46-SERIES
Generic NameWASHER DISINFECTOR
Product CodeMDZ
Date Received2013-07-17
Model Number46-4
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerGETINGE DISINFECTION AB
Manufacturer AddressLJUNGADALSGATAN 11 VAXJO 35115 SW 351 15


Patients

Patient NumberTreatmentOutcomeDate
10 2013-07-17

© 2026 FDA.report