MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 04 report with the FDA on 2013-04-15 for MOBILEHELP CLASSIC IGB-01 manufactured by Integrity Tracking Dba Mobilehelp.
[3641678]
Earl called stating that today his wife fell, pressed button, nothing happened.
Patient Sequence No: 1, Text Type: D, B5
[10916273]
The device did not cause or contribute to the reported injury. The injury occurred as a result of a fall, reported by the subscribers husband. The device is intended to be used to summon help in an emergency. Device was retruded an evaluated, and it was determined that a component failure affected the way the device communicates to the central station. Device is not labeled as a life sustaining device. The intended use indicates that receipt of the alarm is not guaranteed.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 3008101184-2013-00002 |
MDR Report Key | 3266650 |
Report Source | 04 |
Date Received | 2013-04-15 |
Date of Report | 2013-04-10 |
Date of Event | 2013-03-14 |
Date Mfgr Received | 2013-03-15 |
Date Added to Maude | 2013-08-07 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 0 |
Event Location | 0 |
Manufacturer Contact | MR GARY BECKER |
Manufacturer Street | 3701 FAU BLVD STE 300 |
Manufacturer City | BOCA RATON FL 33431 |
Manufacturer Country | US |
Manufacturer Postal | 33431 |
Manufacturer Phone | 5613476255 |
Single Use | 3 |
Remedial Action | OT |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | MOBILEHELP CLASSIC |
Generic Name | PERSONAL EMERGENCY RESPONSE SYSTEM |
Product Code | ILQ |
Date Received | 2013-04-15 |
Returned To Mfg | 2013-03-27 |
Model Number | IGB-01 |
Operator | OTHER |
Device Availability | Y |
Device Age | DA |
Device Eval'ed by Mfgr | Y |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | INTEGRITY TRACKING DBA MOBILEHELP |
Manufacturer Address | 3701 FAU BLVD STE 300 BOCA RATON FL 33431 US 33431 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Hospitalization | 2013-04-15 |