EO GAS STERILIZER

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2013-08-09 for EO GAS STERILIZER manufactured by Steris Mexico, S. De R.l. De C.v..

Event Text Entries

[3635265] The user facility reported an eto leak. No injuries to hospital staff or patients were reported. The department was evacuated and the fire department was dispatched. No procedural delays or cancellations occurred.
Patient Sequence No: 1, Text Type: D, B5


[10889720] The user facility's 3rd party technician detached the exhaust manifold assembly from the eo sterilizer while repairing the sterilizer. The exhaust manifold is shared with a second eo sterilizer, and when the second sterilizer was operated, eo gas leaked from the first sterilizer due to disconnection of the exhaust manifold assembly from the first sterilizer. The user facility requested a steris technician to inspect and assess the incident; the steris technician confirmed that the eo sterilizer was correctly installed and operating to steris specifications. When the 3rd party technician removed the exhaust manifold, he did not follow the lock out/tag out safety procedures, resulting in the user facility operating the second eo sterilizer which should not have been operated while the first unit was under maintenance.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3005899764-2013-00081
MDR Report Key3274828
Report Source06
Date Received2013-08-09
Date of Report2013-08-09
Date of Event2013-07-10
Date Mfgr Received2013-07-11
Date Added to Maude2013-11-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS KATHRYN CADORETTE
Manufacturer Street5960 HEISLEY ROAD
Manufacturer CityMENTOR OH 44060
Manufacturer CountryUS
Manufacturer Postal44060
Manufacturer Phone4403927231
Manufacturer G1STERIS MEXICO, S. DE R.L. DE C.V.
Manufacturer StreetAVENIDA AVANTE 790 PARQUE INDUSTRIAL GUADALUPE
Manufacturer CityGUADALUPE, NUEVO LEON 67190
Manufacturer CountryMX
Manufacturer Postal Code67190
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameEO GAS STERILIZER
Generic NameSTERILIZER
Product CodeFLF
Date Received2013-08-09
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSTERIS MEXICO, S. DE R.L. DE C.V.
Manufacturer AddressAVENIDA AVANTE 790 PARQUE INDUSTRIAL GUADALUPE GUADALUPE, NUEVO LEON 67190 MX 67190


Patients

Patient NumberTreatmentOutcomeDate
10 2013-08-09

© 2024 FDA.report
This site is not affiliated with or endorsed by the FDA.