OLYMPIC COOL-CAP 60010

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2013-08-09 for OLYMPIC COOL-CAP 60010 manufactured by Natus Medical Inc.

Event Text Entries

[15174915] Facility contacted natus medical incorporated to report the cool-cap malfunction. The device failed to perform as designed. The facility contacted natus medical incorporated to report that the cool cap monitor froze during treatment of a pt. The pt was then moved the pt to another cool cap unit and resumed treatment.
Patient Sequence No: 1, Text Type: D, B5


[15544177] The reported device is part of an ongoing field corrective action. "the olympic cool-cap control module has experienced a frozen screen during use. When this occurs the on-screen info remains on display but the system is no longer providing cooling treatment to the pt. The visible clock in the upper right hand corner of the screen stops advancing. " natus medical incorporated contacted the facility to advise of the failure via (b)(4). The hosp facility did not respond to the corrective action notice delivered to them on 12/12/2012. Natus medical incorporated anticipates that there will be no further f/u to this report, as the issue is understood and is currently being corrected. If new info becomes available the f/u will be filed for this incident. No other info is available. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3018859-2013-00006
MDR Report Key3287715
Report Source07
Date Received2013-08-09
Date of Report2013-07-28
Date of Event2013-07-24
Device Manufacturer Date2007-01-01
Date Added to Maude2013-11-04
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer Street5900 FIRST AVE., SOUTH
Manufacturer CitySEATTLE WA 98108
Manufacturer CountryUS
Manufacturer Postal98108
Manufacturer Phone2062685156
Single Use3
Remedial ActionRC
Previous Use Code3
Removal Correction NumberZ-00447-2013
Event Type3
Type of Report3

Device Details

Brand NameOLYMPIC COOL-CAP
Generic NameCOOL-CAP
Product CodeMXM
Date Received2013-08-09
Model Number60010
Catalog Number60010
Lot NumberNA
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerNATUS MEDICAL INC


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2013-08-09

© 2026 FDA.report