RHOTON HOOK 90DEG BLUNT OVERLGTH 7-1/2IN NL3785-010

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2013-08-15 for RHOTON HOOK 90DEG BLUNT OVERLGTH 7-1/2IN NL3785-010 manufactured by Carefusion.

Event Text Entries

[3605124] This complaint came to carefusion via a medwatch report number (b)(4): it was reported that the tip of the nerve hook instrument broke off and fell into the patient's heart during surgery. Attempts to find it with a magnet unsuccessful. X-ray of the patient did not reveal where the tip was at. X-ray done of cell saver and heart bypass filters. The tip was found in the cell saver filter. Comment 7/19/2013: this medwatch came with product number nl3185-010, carefusion rhoton hooks all present with a nl37 prefix. It is believed this was an error on part of the facility and the product is indeed our nl3785-010, this complaint reflects that? Product number. Writer will attempt to confirm with end user.? Additional information received 7/19/2013: it was reported that the device is available for evaluation.? Customer is requesting the device be returned after carefusion is done with the evaluation. The patient underwent an x-ray but no other medical intervention was preformed. The procedure was completed as planned. The facility confirmed that the tip was found in the filter of the cell saver. The facility has agreed to email photos of the device and x-rays.
Patient Sequence No: 1, Text Type: D, B5


[11055030] (b)(4).
Patient Sequence No: 1, Text Type: N, H10


[24218414] (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1923569-2013-00004
MDR Report Key3290188
Report Source05,06
Date Received2013-08-15
Date of Report2013-07-18
Date of Event2013-05-31
Date Mfgr Received2013-07-18
Device Manufacturer Date2003-04-01
Date Added to Maude2013-08-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationRISK MANAGER
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactJILL RITTORNO
Manufacturer Street75 NORTH FAIRWAY DRIVE
Manufacturer CityVERNON HILLS IL 60061
Manufacturer CountryUS
Manufacturer Postal60061
Manufacturer Phone8474737294
Manufacturer G1CAREFUSION 2200 INC (ST. LOUIS)
Manufacturer Street5 SUNNEN DR
Manufacturer CityST. LOUIS 63143
Manufacturer CountryUS
Manufacturer Postal Code63143
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameRHOTON HOOK 90DEG BLUNT OVERLGTH 7-1/2IN
Generic NameINSTRUMENT, MICROSURGICAL
Product CodeGZX
Date Received2013-08-15
Returned To Mfg2013-07-31
Model NumberNL3785-010
Catalog NumberNL3785-010
Lot NumberD03
OperatorPHYSICIAN
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrR
Device Sequence No1
Device Event Key0
ManufacturerCAREFUSION
Manufacturer Address1500 WAUKEGAN RD MCGAW PARK IL 60085 US 60085


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2013-08-15