MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2001-04-25 for ARTICULATING ARM FOR CEILING MOUNT SYSTEM 5787 manufactured by E-z-em, Inc..
[17171849]
Ct technician was tableside prior to a procedure when the ceiling mount for the injector failed. The horizontal portion of the articulating arm snapped approximately 2 inches from the ceiling column causing the injector and arm to fall. The injector/arm stuck the technician from behind on the right shoulder (patient was not hit). The technician is currently undergoing physical therapy as a result of this injury.
Patient Sequence No: 1, Text Type: D, B5
Report Number | 2432460-2001-00006 |
MDR Report Key | 329123 |
Report Source | 05,06 |
Date Received | 2001-04-25 |
Date of Report | 2001-03-26 |
Date of Event | 2001-03-25 |
Date Facility Aware | 2001-03-25 |
Report Date | 2001-03-26 |
Date Reported to Mfgr | 2001-03-26 |
Date Mfgr Received | 2001-03-26 |
Date Added to Maude | 2001-04-30 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 0 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 3 |
Event Location | 3 |
Manufacturer Contact | JUDITH MERITZ, VP |
Manufacturer Street | 717 MAIN ST |
Manufacturer City | WESTBURY NY 11590 |
Manufacturer Country | US |
Manufacturer Postal | 11590 |
Manufacturer Phone | 8005444624 |
Manufacturer G1 | * |
Manufacturer Street | * |
Manufacturer City | * |
Manufacturer Country | * |
Single Use | 3 |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | ARTICULATING ARM FOR CEILING MOUNT SYSTEM |
Generic Name | ACCESSORY FOR PERCUPUMP CT INJECTOR |
Product Code | FIH |
Date Received | 2001-04-25 |
Returned To Mfg | 2001-04-03 |
Model Number | NA |
Catalog Number | 5787 |
Lot Number | NA |
ID Number | NA |
Operator | HEALTH PROFESSIONAL |
Device Availability | R |
Device Age | UNKNOWN |
Device Eval'ed by Mfgr | Y |
Implant Flag | N |
Date Removed | A |
Device Sequence No | 1 |
Device Event Key | 318508 |
Manufacturer | E-Z-EM, INC. |
Manufacturer Address | 113-117 MAGNOLIA AVE. WESTBURY NY 11590 US |
Baseline Brand Name | ARTICULATING ARM FOR CEILING MOUNT SYSTEM |
Baseline Generic Name | CT INJECTOR ACCESSORY |
Baseline Model No | NA |
Baseline Catalog No | 5787 |
Baseline ID | NA |
Baseline Device Family | ACCESSORY FOR PERCUPUMP INJECTOR SYSTEM |
Baseline Shelf Life [Months] | NA |
Baseline PMA Flag | N |
Baseline 510K PMN | Y |
Premarket Notification | K864227 |
Baseline Preamendment | N |
Baseline Transitional | N |
510k Exempt | N |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 1. Required No Informationntervention | 2001-04-25 |