CLIP APPLICATOR MCEN21 110MM [MXI/XOM]

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,06,07 report with the FDA on 2013-08-20 for CLIP APPLICATOR MCEN21 110MM [MXI/XOM] manufactured by Xomed Microfrance Mfg.

Event Text Entries

[3710872] A device (mcen21) was returned to service - repair alleging a "missing adjustment screw".
Patient Sequence No: 1, Text Type: D, B5


[11070344] (b)(4). Results code is mechanical shock problem. Analysis found the mobile jaw was broken. The other jaw presents a significant trace of shock. The highlighted damages are probably the consequence of shocks during use or reprocessing. Note: this mdr is being filed as a result of an internal review of complaints from medtronic? S mxi facility in (b)(6). This review was conducted to resolve several issues discovered in the mxi complaint handling process. Issues resolved include the misclassification of devices returned for repair, as well as gaps in reporting. Mxi capa (b)(4) was opened to address these issues. Mxi internal reference #:(b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9680837-2013-00032
MDR Report Key3299184
Report Source01,06,07
Date Received2013-08-20
Date of Report2013-04-09
Date of Event2013-04-08
Date Mfgr Received2013-04-09
Device Manufacturer Date2002-04-01
Date Added to Maude2013-11-15
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional0
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactJONATHAN SCHLECH
Manufacturer Street6743 SOUTHPOINT DRIVE NORTH
Manufacturer CityJACKSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal32216
Manufacturer Phone9043328355
Manufacturer G1MEDTRONIC XOMED, INC.
Manufacturer Street6743 SOUTHPOINT DRIVE NORTH
Manufacturer CityJACKSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal Code32216
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCLIP APPLICATOR MCEN21 110MM [MXI/XOM]
Generic NameFORCEPS, ENT
Product CodeKAE
Date Received2013-08-20
Returned To Mfg2013-05-07
Model NumberMCEN21
Catalog NumberMCEN21
Lot Number200204MF
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerXOMED MICROFRANCE MFG
Manufacturer AddressSAINT-AUBIN-LE-MONIAL BOURBON-L'ARCHAMBAULT 3160 FR 3160


Patients

Patient NumberTreatmentOutcomeDate
10 2013-08-20