PUNCH 3712038 CUTTING CIRCULAR 3.5MM

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06,07 report with the FDA on 2013-08-23 for PUNCH 3712038 CUTTING CIRCULAR 3.5MM manufactured by Medtronic Xomed, Inc..

Event Text Entries

[3586595] It was reported that the device broke. It was confirmed with the facility that intra-operatively the device stopped working, it would not open or close, and the physician discontinued use. There were no fragments or pieces that broke off. There was no patient impact. When received for analysis it was found that there was a piece that had broken off/detached from the device.
Patient Sequence No: 1, Text Type: D, B5


[11073541] This device is used for therapeutic purposes. (b)(4). Product evaluation: when received for product analysis there was evidence of customer use; wear marks on working surfaces. When compared to the assembly drawing: the inner shaft broke resulting in a 0. 542? Portion of the tip becoming detached. When viewed under magnification, the break point had pitting which is consistent with rust/corrosion. Without the lot number the age of the device cannot be determined. Method code: microscopic inspection. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1045254-2013-00498
MDR Report Key3304699
Report Source05,06,07
Date Received2013-08-23
Date of Report2013-08-06
Date of Event2013-07-17
Date Mfgr Received2013-08-06
Date Added to Maude2013-11-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMICHELLE ALFORD
Manufacturer Street6743 SOUTHPOINT DRIVE NORTH
Manufacturer CityJACKSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal32216
Manufacturer Phone9043328197
Manufacturer G1MEDTRONIC XOMED, INC.
Manufacturer Street6743 SOUTHPOINT DRIVE NORTH
Manufacturer CityJACKSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal Code32216
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NamePUNCH 3712038 CUTTING CIRCULAR 3.5MM
Generic NamePUNCH, NASAL
Product CodeKAY
Date Received2013-08-23
Returned To Mfg2013-07-29
Model Number3712038
Catalog Number3712038
Lot NumberNI
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC XOMED, INC.
Manufacturer Address6743 SOUTHPOINT DRIVE NORTH JACKSONVILLE FL 32216 US 32216


Patients

Patient NumberTreatmentOutcomeDate
10 2013-08-23

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.