L-CLIP APPL/REMVR 90 MM RND HANDLE VAR 45.441

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 06 report with the FDA on 2013-08-23 for L-CLIP APPL/REMVR 90 MM RND HANDLE VAR 45.441 manufactured by Depuy Synthes Power Tools.

Event Text Entries

[16866020] It was reported that a clip device would "not stay in the locked position". It was unknown to the reporter if the device was used in surgery. It was unknown to the reporter if there were any delays in a surgical procedure or if a spare device was available. It was unknown if injuries or medical intervention were reported. The date of the event was unknown but the reporter clarified that the event occurred in 2013. Several attempts have been made to obtain additional information concerning the reported event; however, no additional information has been provided. A supplemental medwatch report will be submitted if further information is received.
Patient Sequence No: 1, Text Type: D, B5


[17046932] The device has not been returned as this is a single-use instrument; therefore, the reported condition cannot be confirmed and/or duplicated. If additional information should become available, a supplemental medwatch report will be sent accordingly.
Patient Sequence No: 1, Text Type: N, H10


[22088010] The actual device was returned for evaluation. Reliability engineering evaluated the device and observed that the device would not stay in the locked position. Therefore, the reported condition was confirmed. The assignable root cause was determined to be due to misuse, abuse and/or user error. If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1045834-2013-24039
MDR Report Key3304758
Report Source06
Date Received2013-08-23
Date of Report2013-07-26
Date Mfgr Received2014-09-05
Date Added to Maude2013-11-21
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMR. RODNEY CRAIN
Manufacturer Street4500 RIVERSIDE DRIVE
Manufacturer CityPALM BEACH GARDENS FL 33410
Manufacturer CountryUS
Manufacturer Postal33410
Manufacturer Phone5616271080
Manufacturer CountryUS
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameL-CLIP APPL/REMVR 90 MM RND HANDLE VAR
Generic NameNI
Product CodeHCI
Date Received2013-08-23
Returned To Mfg2013-08-27
Catalog Number45.441
Lot NumberNI
OperatorHEALTH PROFESSIONAL
Device AvailabilityR
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerDEPUY SYNTHES POWER TOOLS
Manufacturer Address4500 RIVERSIDE DRIVE PALM BEACH GARDENS FL 33410 US 33410


Patients

Patient NumberTreatmentOutcomeDate
10 2013-08-23