DIMENSION VISTA CLINICAL CHEMISTRY SYSTEM DIMENSION VISTA 1500

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2013-08-27 for DIMENSION VISTA CLINICAL CHEMISTRY SYSTEM DIMENSION VISTA 1500 manufactured by Siemens Healthcare Diagnostics Inc..

Event Text Entries

[3844281] A discordant, falsely elevated gentamicin result was obtained on one patient sample on a dimension vista 1500 instrument. The discordant result was reported to the physician(s), who questioned the result. The sample was repeated on the same instrument and resulted lower. The patient was then redrawn and the new sample was run on the same instrument in duplicate. The results obtained on the new sample matched the rerun result from the initial sample. The rerun results were reported to the physician(s). There are no reports of patient intervention or adverse health consequences due to the discordant, falsely elevated gentamicin result.
Patient Sequence No: 1, Text Type: D, B5


[11013218] A siemens global product support (gps) specialist evaluated the instrument data and did not discover an instrument malfunction. There were no instrument errors when the sample was initially processed and the sample had resulted as expected upon repeat testing on the same instrument. The cause of the discordant, falsely elevated gentamicin result on one patient sample is unknown. The instrument is performing according to specifications. No further evaluation of the device is required.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number1226181-2013-00373
MDR Report Key3311172
Report Source05,06
Date Received2013-08-27
Date of Report2013-08-06
Date of Event2013-08-06
Date Mfgr Received2013-08-06
Device Manufacturer Date2010-07-02
Date Added to Maude2013-11-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactCASSANDRA KOCSIS
Manufacturer Street511 BENEDICT AVENUE
Manufacturer CityTARRYTOWN NY 10591
Manufacturer CountryUS
Manufacturer Postal10591
Manufacturer Phone9145242687
Manufacturer G1SIEMENS HEALTHCARE DIAGNOSTICS INC.
Manufacturer Street101 SILVERMINE ROAD
Manufacturer CityBROOKFIELD CT 06804
Manufacturer CountryUS
Manufacturer Postal Code06804
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameDIMENSION VISTA CLINICAL CHEMISTRY SYSTEM
Generic NameCLINICAL CHEMISTRY SYSTEM
Product CodeLCD
Date Received2013-08-27
Model NumberDIMENSION VISTA 1500
OperatorHEALTH PROFESSIONAL
Device AvailabilityN
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerSIEMENS HEALTHCARE DIAGNOSTICS INC.
Manufacturer Address101 SILVERMINE ROAD BROOKFIELD CT 06804 US 06804


Patients

Patient NumberTreatmentOutcomeDate
10 2013-08-27

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.