THINPREP IMAGING SYSTEM

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,07 report with the FDA on 2013-08-20 for THINPREP IMAGING SYSTEM manufactured by Hologic Inc..

Event Text Entries

[16433711] Lab personnel reported a case with observed abnormal cells that were not located in the 22 fields of view (fov) selected by the imager. Cytology application support specialist (cas) reviewed slide on the review scope as an unk mixed with other slides. Case presented had no triggers in the 22 fovs to prompt an autoscan. Full review of slide during qc showed abnormal cells outside the fovs. Cas reviewed this case and discussed with lab personnel. Cas agreed there were no changes seen in 22 fov to prompt an autoscan. There were two side by side cells each contained large koilocytes with binucleated hyperchromatic nuclei. The overall appearance of the slide was clean. Pt was a (b)(6) female, day 17 lmp; specimen was cervical. Abnormal cells were discovered by the lab when tech was scanning around after looking at the 22 fobs. Laboratory's interpretation of cells outside fov was lgsil-hpv. Cas interpretation of cells outside fov was lgsil-hpv. Site will be monitored to determine if expected occurrences (as determined by the pivotal study) are exceeded. Cytotechnologists on site will continued to hold questionable cases for cas review on next visit.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1222780-2013-00156
MDR Report Key3313765
Report Source05,07
Date Received2013-08-20
Date of Report2013-08-01
Date of Event2013-08-01
Date Mfgr Received2013-08-01
Date Added to Maude2013-09-05
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactRANDALL COVILL, MANAGER
Manufacturer Street250 CAMPUS DRIVE
Manufacturer CityMARLBOROUGH MA 01752
Manufacturer CountryUS
Manufacturer Postal01752
Manufacturer Phone5082638853
Single Use0
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTHINPREP IMAGING SYSTEM
Generic NameAUTOMATED MICROSCOPE FOR CYTOLOGY
Product CodeMNM
Date Received2013-08-20
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerHOLOGIC INC.
Manufacturer AddressMARLBOROUGH MA US


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2013-08-20

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.