[3699044] The user facility reported their unit was leaking water near electrical wires. The leak was approximated to be one gallon of water. No injuries or procedural delays/cancellations were reported.
Patient Sequence No: 1, Text Type: D, B5


[10921906] A steris service technician arrived at the facility and found the copper inlet water pipe to the water temperature booster tank leaking water. The leak caused water to contact the electrical wires to the inlet solenoid valves. The technician replaced the copper piping and cleaned the water level switches. The unit was installed in november of 1997 and has not been under a steris service contract since 2012. The replacement of the copper piping is attributed to the age of the component.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9680353-2013-00117
MDR Report Key3325517
Report Source06
Date Received2013-09-04
Date of Report2013-09-04
Date of Event2013-08-05
Date Mfgr Received2013-08-05
Date Added to Maude2013-12-03
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationOTHER HEALTH CARE PROFESSIONAL
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactMS KATHRYN CADORETTE
Manufacturer Street5960 HEISLEY ROAD
Manufacturer CityMENTOR OH 44060
Manufacturer CountryUS
Manufacturer Postal44060
Manufacturer Phone4403927231
Manufacturer G1STERIS CANADA CORPORATION
Manufacturer Street490, ARMAND-PARIS
Manufacturer CityQUEBEC, QUEBEC GIC8A3
Manufacturer CountryCA
Manufacturer Postal CodeGIC 8A3
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand Name120L CART WASHER
Generic NameCART WASHER
Product CodeFLH
Date Received2013-09-04
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerSTERIS CANADA CORPORATION
Manufacturer Address490 ARMAND-PARIS QUEBEC, QUEBEC GIC8A3 CA GIC 8A3


Patients

Patient NumberTreatmentOutcomeDate
10 2013-09-04

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.