TCM400 TCM400 BASE UNIT 391-880

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,05 report with the FDA on 2013-09-03 for TCM400 TCM400 BASE UNIT 391-880 manufactured by Radiometer Medical Aps.

Event Text Entries

[17342173] Thursday, (b)(6) 2013, an operator switched off the mains switch of the analyzer and disconnected the line cord from the power outlet. During transport the operator wanted to make sure that it was the correct analyzer to ship for repair. While checking the serial number on the back of the monitor the operator's left forearm got in contact with the analyzer housing/ chassis. At the same time, he also got in contact with the male connector of the line cord and received an electrical shock. This happened 2 to 3 minutes after he disconnected the monitor from mains. The operator had visible marks on his left forearm and left ring finger from the electrical current flow, and was admitted for long term ecg monitoring for 24 hours. There was no further outcome for the patient except an occasional prickling in the affected arm.
Patient Sequence No: 1, Text Type: D, B5


[17549432] The monitor has been returned for investigation. The error could not be re-constructed, but it was confirmed that the current event was not related to fan (b)(4) (3002807968-10/19/12-003-c initiated 2012-11-27). Further tests will be performed on the monitor.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number3002807968-2013-00014
MDR Report Key3330542
Report Source01,05
Date Received2013-09-03
Date of Report2013-08-22
Date of Event2013-08-15
Date Mfgr Received2013-08-22
Date Added to Maude2013-09-09
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactLASSE MOLLER, SPECIALIST
Manufacturer StreetAKANDEVEJ 21
Manufacturer CityBRONSHOJ DK-2700
Manufacturer CountryDA
Manufacturer PostalDK-2700
Manufacturer Phone8273436
Single Use0
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameTCM400
Generic NameTCM400
Product CodeLPP
Date Received2013-09-03
Returned To Mfg2013-08-26
Model NumberTCM400 BASE UNIT
Catalog Number391-880
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device AgeDA
Device Eval'ed by Mfgr*
Device Sequence No1
Device Event Key0
ManufacturerRADIOMETER MEDICAL APS
Manufacturer AddressAKANDEVEJ 21 BRONSHOJ DK-2700 DA DK-2700


Patients

Patient NumberTreatmentOutcomeDate
101. Hospitalization 2013-09-03

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.