MIDLAND BARIATRIC STAND -IN TABLE 554512E *

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed with the FDA on 2013-08-11 for MIDLAND BARIATRIC STAND -IN TABLE 554512E * manufactured by Patterson Medical.

Event Text Entries

[3866843] Pt was seated in wheelchair with 2 therapists present attempting to stand pt in bariatric standing frame. The frame was being lowered to the ground to lock into place and was lowered onto patient's left big toe. Pt reported pain and the stand was immediately lifted from toe. Therapist saw blood on sock. Two therapists present removed long stocking sock and grip sock, cleaned toe with wet cloth. Toe did not appear to be broken or severely injured. Follow-up investigation: witness had stated that this has happened before one time that he could remember. From the details related, the normal protocol is to have the stand in the lowered position before patient is attached, then secure and lift the patient. Apparently during this lift some readjustments were needed including moving the position of the stand. So the frame was raised and lowered again while patient was attached. This is when the injury occurred. Some discussion on design improvement was brought up. A couple of suggestions were to have the knee pad made adjustable to accommodate different shaped patients and some kind of guard to prevent toes from going under the frame. Maude database checked and one similar report was found. Recommendations from the manufacturer were apply a warning sticker on the machine advising to make sure lift is lowered position before lifting patient. The device in question had 2 of these on it. One by the handle that controls the raising and lowering of the frame and the other was on the back side of the stand. The ot unit only has one of these devices and it is needed for this patient so it was put back in use. Ot supervisor is going to work on additional labeling and education of staff to not raise the frame once a patient is secured to the stand. From conversation from staff they felt that this is both a use problem and a design issue. ====================== manufacturer response for standing frame, bariatric standing frame (per site reporter). ====================== recommendations from the manufacturer were apply a warning sticker on the machine advising to make sure lift is lowered position before lifting patient. The device in question had 2 of these on them.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number3337685
MDR Report Key3337685
Date Received2013-08-11
Date of Report2013-06-12
Date of Event2013-06-12
Report Date2013-06-12
Date Reported to FDA2013-08-11
Date Reported to Mfgr2013-09-10
Date Added to Maude2013-09-10
Event Key0
Report Source CodeUser Facility report
Manufacturer LinkN
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location3
Single Use0
Previous Use Code0
Event Type3
Type of Report3

Device Details

Brand NameMIDLAND BARIATRIC STAND -IN TABLE
Generic NameWHEELCHAIR, STANDUP
Product CodeIPL
Date Received2013-08-11
Model Number554512E
Catalog Number*
Lot Number*
ID Number*
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Sequence No1
Device Event Key0
ManufacturerPATTERSON MEDICAL
Manufacturer Address1031 MENDOTA HEIGHTS RD SAINT PAUL MN 55120 US 55120


Patients

Patient NumberTreatmentOutcomeDate
10 2013-08-11