CUP FORCEPS MCEN62A1 45 DEG N1 JANKOWSKY

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 01,07 report with the FDA on 2013-08-29 for CUP FORCEPS MCEN62A1 45 DEG N1 JANKOWSKY manufactured by Medtronic Xomed Instrumentation S.a.s..

Event Text Entries

[3722566] The device (part # mcen62a1) was returned to mxi service and repair.
Patient Sequence No: 1, Text Type: D, B5


[11089726] The device (part # mcen62a1) was evaluated and indicated that fixed jaw was broken at the pin connection. Too much stress and lack of material around the pin. (b)(4). Note: this mdr is being filed as a result of an internal review of complaint from medtronic's mxi facility in (b)(4). This review was conducted to resolve several issues discovered in the mxi complaint handling process. Issues resolved include the misclassification of devices returned for repair, as well as gaps in reporting. (b)(4).
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9680837-2013-00163
MDR Report Key3338062
Report Source01,07
Date Received2013-08-29
Date of Report2011-05-17
Date of Event2011-05-17
Date Mfgr Received2011-05-17
Device Manufacturer Date2011-01-01
Date Added to Maude2013-12-02
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Reporter OccupationBIOMEDICAL ENGINEER
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location0
Manufacturer ContactJONATHAN SCLECH
Manufacturer Street6743 SOUTHOINT DT N
Manufacturer CityJACKSONVILLE FL 32216
Manufacturer CountryUS
Manufacturer Postal32216
Manufacturer Phone9043328355
Manufacturer G1MEDTRONIC XOMED INSTRUMENTATION S.A.S.
Manufacturer StreetLE PAVILLON
Manufacturer CitySAINT-AUBIN-LE-MONIAL 03160
Manufacturer CountryFR
Manufacturer Postal Code03160
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameCUP FORCEPS MCEN62A1 45 DEG N1 JANKOWSKY
Product CodeKAE
Date Received2013-08-29
Returned To Mfg2011-05-17
Model NumberMCEN62A1
Lot Number201101MF1
Device AvailabilityY
Device AgeDA
Device Eval'ed by MfgrY
Device Sequence No1
Device Event Key0
ManufacturerMEDTRONIC XOMED INSTRUMENTATION S.A.S.
Manufacturer AddressLE PAVILLON SAINT-AUBIN-LE-MONIAL 03160 FR 03160


Patients

Patient NumberTreatmentOutcomeDate
10 2013-08-29

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.