ADULT LUMBAR PUNCTURE TRAY 20X3 1/2 4301C

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 05,06 report with the FDA on 2001-05-24 for ADULT LUMBAR PUNCTURE TRAY 20X3 1/2 4301C manufactured by Allegiance Healthcare Corp..

Event Text Entries

[189336] Spoke with asst. Nurse mgr on 4/30/01, who stated on the day of the event, a pt had a lumbar puncture performed. The stopcock leaked and they were unable to obtain a pressure reading. They opened another kit and had the same experience. They scheduled a follow up lumbar puncture for the following tuesday. At that time, the child had the third lumbar puncture, and they were able to obtain a pressure reading and noted that the extension tubing was leaking.
Patient Sequence No: 1, Text Type: D, B5


MAUDE Entry Details

Report Number1423507-2001-00162
MDR Report Key334639
Report Source05,06
Date Received2001-05-24
Date of Report2001-05-24
Date of Event2001-04-20
Date Mfgr Received2001-04-27
Device Manufacturer Date2000-11-01
Date Added to Maude2001-05-29
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag0
Health Professional3
Initial Report to FDA3
Report to FDA3
Event Location0
Manufacturer ContactPATRICIA SHARPE-GREGG
Manufacturer Street1500 WAUKEGAN ROAD
Manufacturer CityMCGAW PARK IL 60085
Manufacturer CountryUS
Manufacturer Postal60085
Manufacturer Phone8475784148
Manufacturer G1*
Manufacturer Street*
Manufacturer City*
Manufacturer Country*
Single Use3
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameADULT LUMBAR PUNCTURE TRAY 20X3 1/2
Generic NamePROCEDURE TRAY
Product CodeIQE
Date Received2001-05-24
Model Number4301C
Catalog Number4301C
Lot NumberL0P049
ID NumberUNK
OperatorHEALTH PROFESSIONAL
Device AvailabilityY
Device Eval'ed by MfgrY
Implant FlagN
Date RemovedU
Device Sequence No1
Device Event Key323984
ManufacturerALLEGIANCE HEALTHCARE CORP.
Manufacturer Address1500 WAUKEGAN RD. MCGAW PARK IL 60085 US
Baseline Brand NameLUMBAR PUNCTURE TRAY
Baseline Generic NameLUMBAR PUNCTURE
Baseline Model No4301C
Baseline Catalog No4301C
Baseline IDNA
Baseline Device FamilyPROCEDURE TRAY
Baseline Shelf Life Contained*
Baseline Shelf Life [Months]*
Baseline PMA FlagN
Baseline 510K PMNN
Baseline PreamendmentN
Baseline TransitionalN
510k ExemptN


Patients

Patient NumberTreatmentOutcomeDate
101. Other 2001-05-24