MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2013-09-13 for GETINGE 46-SERIES 46-4 manufactured by Getinge Disinfection Ab.
[20109057]
Smoke/fire was observed coming from a getinge model 46-4 washer disinfector during use. The washer was approx 35-40 minutes into a 70 minute cycle when staff observed smoke/fire. The user's turned off the power to the unit and sprayed with a dry powder extinguisher.
Patient Sequence No: 1, Text Type: D, B5
[20482096]
Device was evaluated remotely (not returned to manufacturer) via investigation reports and photographs from getinge (b)(4). , a service unit located in (b)(4). It was concluded that the overheat protection system failed/delayed response to an overheat condition. A voluntary device correction of getinge model 46-4 washer disinfectors was reported to u. S. Fda on (b)(4) 2013 (recall number not yet assigned). This correction was initiated to address two conditions. A potential trigger for overheating; (root cause) - inadequate overheat protection system component. A component (shunt trip) selected for the overheat protection system does not meet the required parameters (control voltage) where/when overheating occurs. The correction entails - inspecting the setting of a software function that can be utilized during servicing of a unit to skip wash phases. This is a password protected function but if left in the enabled state, a user can inadvertently activate the heating element with no water in the chamber. The device correction action includes inspecting that this function is not enabled. The installation of a newer design of overheat protection system that utilizes a heating element with integral thermal fuses that meet the required overheat parameters. Recall number not yet assigned.
Patient Sequence No: 1, Text Type: N, H10
Report Number | 9616031-2013-00004 |
MDR Report Key | 3368406 |
Report Source | 07 |
Date Received | 2013-09-13 |
Date of Report | 2013-08-14 |
Date of Event | 2013-08-13 |
Date Mfgr Received | 2013-08-14 |
Date Added to Maude | 2013-09-25 |
Event Key | 0 |
Report Source Code | Manufacturer report |
Manufacturer Link | Y |
Number of Patients in Event | 0 |
Adverse Event Flag | 3 |
Product Problem Flag | 3 |
Reprocessed and Reused Flag | 3 |
Health Professional | 3 |
Initial Report to FDA | 3 |
Report to FDA | 0 |
Event Location | 3 |
Manufacturer Contact | ANN WHEELER, US AGENT |
Manufacturer Street | 1777 E HENRIETTA RD |
Manufacturer City | ROCHESTER NY 14623 |
Manufacturer Country | US |
Manufacturer Postal | 14623 |
Manufacturer Phone | 5852725036 |
Manufacturer G1 | GETINGE DISINFECTION AB |
Manufacturer Street | LJUNGADALSGATAN 11 |
Manufacturer City | VAXJO 35115 |
Manufacturer Country | SW |
Manufacturer Postal Code | 351 15 |
Single Use | 3 |
Remedial Action | RP |
Previous Use Code | 3 |
Event Type | 3 |
Type of Report | 3 |
Brand Name | GETINGE 46-SERIES |
Generic Name | WASHER DISINFECTOR |
Product Code | MDZ |
Date Received | 2013-09-13 |
Model Number | 46-4 |
Operator | OTHER |
Device Availability | N |
Device Age | DA |
Device Eval'ed by Mfgr | N |
Device Sequence No | 1 |
Device Event Key | 0 |
Manufacturer | GETINGE DISINFECTION AB |
Manufacturer Address | LJUNGADALSGATAN 11 VAXJO 35115 SW 351 15 |
Patient Number | Treatment | Outcome | Date |
---|---|---|---|
1 | 0 | 2013-09-13 |