GETINGE 46-SERIES 46-4

MAUDE Adverse Event Report

MAUDE data represents reports of adverse events involving medical devices. This maude entry was filed from a 07 report with the FDA on 2013-09-13 for GETINGE 46-SERIES 46-4 manufactured by Getinge Disinfection Ab.

Event Text Entries

[20109057] Smoke/fire was observed coming from a getinge model 46-4 washer disinfector during use. The washer was approx 35-40 minutes into a 70 minute cycle when staff observed smoke/fire. The user's turned off the power to the unit and sprayed with a dry powder extinguisher.
Patient Sequence No: 1, Text Type: D, B5


[20482096] Device was evaluated remotely (not returned to manufacturer) via investigation reports and photographs from getinge (b)(4). , a service unit located in (b)(4). It was concluded that the overheat protection system failed/delayed response to an overheat condition. A voluntary device correction of getinge model 46-4 washer disinfectors was reported to u. S. Fda on (b)(4) 2013 (recall number not yet assigned). This correction was initiated to address two conditions. A potential trigger for overheating; (root cause) - inadequate overheat protection system component. A component (shunt trip) selected for the overheat protection system does not meet the required parameters (control voltage) where/when overheating occurs. The correction entails - inspecting the setting of a software function that can be utilized during servicing of a unit to skip wash phases. This is a password protected function but if left in the enabled state, a user can inadvertently activate the heating element with no water in the chamber. The device correction action includes inspecting that this function is not enabled. The installation of a newer design of overheat protection system that utilizes a heating element with integral thermal fuses that meet the required overheat parameters. Recall number not yet assigned.
Patient Sequence No: 1, Text Type: N, H10


MAUDE Entry Details

Report Number9616031-2013-00004
MDR Report Key3368406
Report Source07
Date Received2013-09-13
Date of Report2013-08-14
Date of Event2013-08-13
Date Mfgr Received2013-08-14
Date Added to Maude2013-09-25
Event Key0
Report Source CodeManufacturer report
Manufacturer LinkY
Number of Patients in Event0
Adverse Event Flag3
Product Problem Flag3
Reprocessed and Reused Flag3
Health Professional3
Initial Report to FDA3
Report to FDA0
Event Location3
Manufacturer ContactANN WHEELER, US AGENT
Manufacturer Street1777 E HENRIETTA RD
Manufacturer CityROCHESTER NY 14623
Manufacturer CountryUS
Manufacturer Postal14623
Manufacturer Phone5852725036
Manufacturer G1GETINGE DISINFECTION AB
Manufacturer StreetLJUNGADALSGATAN 11
Manufacturer CityVAXJO 35115
Manufacturer CountrySW
Manufacturer Postal Code351 15
Single Use3
Remedial ActionRP
Previous Use Code3
Event Type3
Type of Report3

Device Details

Brand NameGETINGE 46-SERIES
Generic NameWASHER DISINFECTOR
Product CodeMDZ
Date Received2013-09-13
Model Number46-4
OperatorOTHER
Device AvailabilityN
Device AgeDA
Device Eval'ed by MfgrN
Device Sequence No1
Device Event Key0
ManufacturerGETINGE DISINFECTION AB
Manufacturer AddressLJUNGADALSGATAN 11 VAXJO 35115 SW 351 15


Patients

Patient NumberTreatmentOutcomeDate
10 2013-09-13

© 2026 FDA.report
This site is not affiliated with or endorsed by the FDA.